Drager Air-Shields Resuscitaire RW Features
The Drager Air-Shields Resuscitaire RW infant warmer combines an effective warming therapy platform with the components you need for clinical emergency resuscitation. The Drager Resuscitaire offers a blender, suction, flowmeter, and positive pressure mounted on the bed for easy accessibility. It features a Silica Quartz Heater that provides complete, even, and quick heating. A parabolic reflector helps distribute the heat evenly over the entire mattress area.
- Exceptional access from either side.
- Configure to your department’s needs.
- Easily access suction.
- Efficiently adjust mattress position.
- The head of the warmer module can swivel up to 90 degrees.
- Automatic Pre-Warm Mode allows long-term, controlled warming of the mattress with no nuisance alarms.
Drager Air-Shields Resuscitaire RW Specifications
Physical Attributes (Without Options / Accessories)
- Height: 72 in (183 cm)
- Width: 26.5 in (68 cm)
- Depth: 44 in (112 cm)
- Weight: (without options /accessories) 220–280 lbs (100–127 kg)
Mattress
- Height From Floor: 39.4 in (100 cm)
- Width: 21.0 in (53.3 cm)
- Length: 26.0 in (66.0 cm)
- Thickness: 1.0 in (2.54 cm)
- Maximum Capacity: 15 lbs (4.54 kg)
Maximum Power Requirements
- Model 120V: 120 V, 50–60 Hz, 750 W
- Model 220/240V: 220–240 V, 50–60 Hz, 750 W )
LED Examination Light
- Illumination: > 0.11 lumens / cm2 (100 ft. Candles)
Controller Heating Alarms
- Check patient System alerts for 5 minutes in Manual Mode every 30 seconds after 10 minutes
- Baby temperature ±1° C from Set Point
- High-temperature Skin Temp. 39.0 ±0.2 °C
- Probe Short or open circuit / No probe
- System fail Indicates System Fail
- Power fail AC Power Interruption
Alarm Silence / Reset Intervals
- Check patient Resets clock for 10 minutes in manual mode
- Baby temperature 10 minutes
- High temperature 2 minutes
- Procedural silence Presilences baby temp alarm for 5 minutes
Set Temperature / Heater Output
- Set Temp Range: 34–38 °C (>37.0° override)
- Heater Output: 0–100%, 10% increments
- Skin Temp Display: 18–43 °C
- Accuracy: ±0.2 °C (31 to 37° C)
Resuscitation Module
- Wall Supply Pressure: 40–75 psi (275–517 kPa)
- Cylinder Pressure: 2,900 psi max (19 994 kPa)
Patient Gas Supply
- Flow Control Range: 0–15 LPM
- Airway Pressure Limit: Adjustable 0–50 cm H2O (0–4.9 kPa); Internal preset limit 60 cm H2O (5.9 kPa)