The Drager SCIO gas measurement module is a free-standing unit that samples breathing gases from adults to pediatric patients in non-, partial- and total rebreathing systems. The Drager SCIO continuously measures the content of CO2, N2O, O2, and one of the anesthetic agents Halothane, Isoflurane, Enflurane, Sevoflurane, and Desflurane in any mixture. It communicates real-time and derived gas information to the host system.

  • Uses infrared light to continuously measure the sample contents.
  • 150 ml/min sample flow rate.
  • Connects to the SC 7000/8000/9000 XL monitors.
  • Self-zeroing and does not need calibration by clinical staff.
  • O2 cell in the unit does not require routine calibration, however, the user may choose to perform a 1-point or 2-point calibration.

Drager SCIO Specifications

Dimensions

  • Height: 4.8” (12.2 cm)
  • Width: 8.7” (22.2 cm)
  • Depth: 11.7” (30 cm)
  • Weight: 6.6 lb (3 kg)

General

  • Mode of operation: Continuous
  • Power: from a specified power supply
  • Sound Pressure level: = 45 dB(A)
  • Air Ingression, leakage: < 45 ml during zeroing, < 10 ml/min leakage
  • Sample Flow rate: 150 ml/min. ± 20 ml/min.

MultiGas

  • Parameter display: etCO2*, iCO2*, RRc*, et O2, iO2, O2, N2O, etHAL, iHAL, etISO, iISO, etENF, iENF, etSEV, iSEV, etDES, iDES
  • Agents measured: Halothane, Isoflurane, Enflurane, Sevoflurane, Desflurane
  • Display Values
    • Agents, CO2, O2: Inspired and expired concentrations (%), plus trend, and waveform
    • N2O: Inspired and expired concentrations (%)
  • Measuring methods
    • CO2, Agents, N20: Infrared
    • O2: Paramagnetic sensor
  • Display Range
    • CO2: 0% to 10% (resolution 0.1%)
    • O2: 0% to 100% (resolution 1.0%)
    • N2O: 0% to 100% (resolution 1.0%)
    • Agents:
      • Halothane: 0% to 10% (resolution 0.1%)
      • Isoflurane: 0% to 10% (resolution 0.1%)
      • Enflurane: 0% to 10% (resolution 0.1%)
      • Sevoflurane: 0% to 11% (resolution 0.1%)
      • Desflurane: 0% to 24% (resolution 0.1%)

Accuracy

  • Full Accuracy after warmup complete – Agents: (when RR = 40)
    • Halothane (up to 8.5%vol): ±(0.15 vol % + 15% rel.)
    • Isoflurane (up to 8.5%vol): ±(0.15 vol % + 15% rel.)
    • Enflurane (up to 10%vol): ±(0.15 vol % + 15% rel.)
    • Sevoflurane (up to 10%vol):±(0.15 vol % + 15% rel.)
    • Desflurane (up to 20%vol): ±(0.15 vol % + 15% rel.)
  • Full Accuracy after warmup complete – N2O: (when RR = 40) ±(2 vol % + 8% rel.)
  • Full Accuracy after warmup complete – O2: (when RR = 40) ±3 vol %
  • Full Accuracy after warmup complete – CO2: (when RR = 40) ±0.5 vol % or ±12% rel., whichever is greater

Rise Time (10% to 90% at
150 ml/min. flow rate)

  • CO2: <500 ms
  • O2: <650 ms
  • N2O: <500 ms
  • Agents: <500 ms

Delay Time with water
trap and 2.5m sample line:

  • < 4 sec.

Respiration Rate (RRc):

  • Display: Respiration Rate (rpm)
  • Measuring Range:0 to 90 rpm
  • Resolution: 1 rpm
  • Accuracy: 0 to 60 rpm: ±1 rpm; >60 rpm: not specified

Refurbished Drager SCIO Agent Monitor

Soma Tech Intl offers the Drager SCIO agent monitors up to 50% below OEM prices with the same service and warranty as new. Soma technically and cosmetically refurbishes these electrosurgical units back to the original engineering manufacturer’s specifications. Using highly skilled, trained, and certified in-house biomedical engineers these units are fully tested, disassembled, and if necessary, parts are replaced. Once all the parts of the Drager SCIO agent Monitor are in working order, the electrosurgical unit is calibrated back to OEM specifications, to make sure it works exactly the way it did when it originally left the manufacturer. Soma Tech Intl not only refurbished the product, but each unit also goes through a Special cosmetic restoration process. The Drager SCIO is cleaned and minor scratches and dentures are repaired, and decals are replaced if necessary.

Soma Tech Intl also purchases used and pre-owned Agent Monitor. If you or your facility wants to sell your Drager SCIO, Soma Tech Intl has an experienced purchasing department that makes the process headache and hassle-free. When selling your agent monitor to Soma, there is no need for a middle man, it is a direct transaction to Soma which ensures that you get top dollar for your medical equipment. Soma Tech Intl not only buys agent monitors but a wide variety of different medical equipment as well if your medical facility is looking to free up space. If you have medical equipment you would like to sell to Soma, send us a message here!

Banner

Questions about our Anesthesia Monitors and Agent Monitors?

Soma Tech Intl offers a wide range of Anesthesia Monitors and Agent Monitors. If you have any questions about any of our monitors or need a quick quote, call us and one of our knowledgeable sales representatives will help you.

Looking to sell your Anesthesia Monitors and Agent Monitors? Soma ensures your facility gets top dollar for your equipment and also gives the option for trade-in credit.

1-800-GET-SOMA
Contact Us