Abbott Plum XLD
Infusion Pump
Features:
Macro Mode
Micro Mode
Specifications
Dimensions ~ 8.25”h x 7.5”w x 8.75”d (including
pole clamp)
Weight ~ 7.75 lbs (with battery)
Casing High-impact plastic
Power Requirements 100-130 Vac, 50 to 60 Hz, less than 35 W
Fuses 0.5 A, 250 V. slo-blo
Battery Sealed lead-acid rechargeable 8 V battery, internal to
infusion system.
Battery Operation A fully charged new battery will provide eight
hours of operation at 125 ml/hr, or 1000 ml total volume delivered, whichever
occurs first.
Recharge The battery charges whenever the infusion system is
connected to AC (mains) power. If the infusion system is operating at 125 ml/hr,
a full recharge takes approximately 16 hours. If the infusion system is turned
to OFF/CHARGE, recharge takes approximately eight hours.
Delivery Rate Range
Micro Mode .1 to 99.9 ml/hr
Macro Mode 1 to 999 ml/hr (in 1 ml increments)
KVO 1 ml/hr
Dose Limit Range 1 to 9999 ml/hr (in 1 ml increments)
Occlusion Range 10 psig (+5, -2 psig)