Zoll R Series – ALS/BLS Zoll Models

Quotes in less than 24 hours Download Specifications

R Series Features

The Zoll R Series is an ALS defibrillator, BLS defibrillator, and monitor that can provide CPR, defibrillation, and pacing therapies. The Zoll R is not just known as being a defibrillator it offers 3-lead ECG, pacing, end-tide Co2, Non-invasive Blood Pressure (NIBP), and SpO2 Monitoring. With wireless connectivity, the R Series defibrillator can give health providers a quick overall snapshot of the patient’s health via a PC or mobile device. With a color display screen, you can view the CPR dashboard which can show depth, rate, and visual indicators of compression releases. Request a quotation for a price or a certain configuration/package. Soma also offers crash carts that can be used with this series defibrillator. The previous model of the Zoll Defibrillator is the Zoll E Series.

  • Always Code-Ready
  • Remote viewing and administration from a PC or mobile device
  • History of readiness logs for the Joint Commission
  • Readiness check includes defib, pacing, and ECG modules; AC power; battery; 30-joule shock test; electrodes attached and not expired
  • Customizable dashboard

Zoll Defibrillator R Series Specifications

Dimensions

  • Height: 8.2” (20.8 cm)
  • Width: 10.5” (26.7 cm)
  • Depth: 12.5” (31.7 cm)
  • Weight: 15.2 lbs (6.9 kg)

Power

  • Voltage: 100-120 V AC
  • Frequency: 50/60 Hz
  • Battery: Rechargeable Lithium-Ion battery pack

Defibrillation

  • Waveform: Rectilinear Biphasic
  • Patient Impedance Range: 15 to 300 ohms
  • Energy Selections: 1 to 10, 15, 20, 30, 50, 75, 100, 120, 150, and 200 joules selected using controls on the front of the defibrillator or sternum paddles. (Note: When using appropriate pediatric resuscitation electrodes, the 75-joule setting is replaced by 70- and 85-joule settings.)
  • Smart Step Energy Levels: Automatically escalates energy through a configured adult or pediatric protocol
  • Energy Display: Shown on the monitor for both selected and delivered
  • Charge Time: Less than 7 seconds with a new, fully charged battery (first 15 charges to 200 joules); longer charge times may result in a depleted or older battery
  • Synchronized Mode: Synchronizes defibrillator pulse to patient’s R wave. “SYNC” message displayed on monitor and markers shown on both monitor and recorded ECG
  • Charge Controls: Control from the front of the defibrillator or apex paddle
  • Paddles: External apex/sternum paddles; adult plates slide off to expose pediatric electrode surface
  • Code Readiness Testing (OneStep): Verifies defibrillator hardware, therapy delivery cable (with both paddles and electrodes), electrode condition, and expiration (with select OneStep electrodes) without the need for a separate test fixture.

ECG Monitoring

  • Patients Connection: 3-lead ECG cable, 5-lead ECG cable, paddles, or hands-free electrodes; selectable by the front panel switch
  • Input Protection: Fully defibrillator protected. Circuits designed to prevent distortion of ECG signal by pacer pulse
  • Implanted Pacemaker Spike Display: Circuits designed to detect most implanted pacemaker spikes and display a marker on the ECG trace
  • Bandwidth: 0.5 to 21 Hz (-3dB) standard; 0.05 to 150 Hz diagnostic with configurable options of 0.5 Hz to 40 Hz or 1 Hz to 21 Hz
  • Lead Selection: I, II, III aVR, aVL, aVF, V, P1, P2, P3 with OneStep Pacing electrode
  • ECG Size: 0.5, 1.0, 1.5, 2.0, or 3.0 cm/mV display on the monitor
  • Heart Rate Display: 0 to 300 bpm ±5%
  • Heart Rate Alarm: User selectable for tachycardia at 60 to 280 bpm; for bradycardia at 20 to 100 bpm. On/Off status displayed on the screen.

Display

  • Screen Type: Color, VGA liquid crystal display (LCD)
  • Screen Size: 6.5 inches (16.5 cm) diagonally
  • Sweep Speed: 25 mm/sec
  • Viewing Time: 5 seconds with the standard display format
  • Channels: 3
  • Information: Heart Rate, Leads/Pads, Alarm On/Off, Selected Energy, Delivered Energy, User Prompts and Warnings, Code Readiness Test Results, SpO2, Non-invasive Blood Pressure, EtCO2, Pacer Functions, Code Markers, CPR Dashboard

Battery Packs

  • Type: 10.8 V (nominal) rechargeable lithium-ion
  • Capacity: 5.8 amp hours
  • Weight: 1.7 lb (0.77 kg)
  • Recharge Time: 5 hours or less with an integral charger
  • Operating Time: >4 hours of continuous ECG monitoring; 100 maximal energy (200 joules) discharges; 3.5 hours of continuous ECG monitoring and pacing at 60 mA, 80 ppm

Recorder

  • Technology: 90 mm thermal array; 80 mm grid width
  • Speed: 25 mm/sec, 6-second delay
  • Zoll Printing Modes: Manual or automatic
  • Annotations: Time, date, defibrillation energy, patient impedance, heart rate, pacer output, QRS synchronization marker, ECG size, ECG lead, alarm, defibrillator test results, analyze ECG, ECG bandwidth

External Pacing

  • Type: VVI demand; asynchronous (fixed rate) when used without ECG leads or in asynchronous (ASYNC) pacing mode
  • Pulse: Rectilinear, constant current: 40 ms ±2 ms; variable 0 to 140 mA ±5% or 5 mA, whichever is greater. The rate is variable from 30 to 180 ppm ±1.5%
  • Output Protection: Fully defibrillator protected and isolated
  • Zoll Pacing: Eliminates the need to connect separate ECG leads and or a cable when used in conjunction with official electrodes.

Pulse Oximetry with Masimo SET Technology

  • Saturation Range: 1-100% (%SpO2 ) with a resolution of 1%
  • Pulse Rate Range: 25-240 ppm with a resolution of 1 ppm
  • Saturation Accuracy: Non-motion conditions ±2% for adults/pediatrics; ±3% for neonates. During motion ±3% for all patients
  • Pulse Rate Accuracy: Non-motion conditions ±3 ppm. During motion ±5 ppm.

Mainstream CO2 Capnostat 5 Sensor

  • Principle of Operation: Nondispersive infrared (NDIR) single-beam optics, dual-wavelength, no moving parts
  • Warm-up Time: Full specifications within 2 minutes at an ambient temperature of 25°C. Capnogram in 20 seconds
  • Environmental: Zoll Operating Temperature: 0°C to 45°C, Storage and Shipping Temperature: -40°C to 70°C.

Sidestream CO2 LoFlo Sensor

  • R Series Principle of Operation: Nondispersive infrared (NDIR) single-beam optics, dual-wavelength, no moving parts
  • R Series Warm-Up Time: Full specifications within 2 minutes at an ambient temperature of 25ºC. Capnogram in 20 seconds
  • R Series Environmental: Operating Temperature: 0°C to 40°C, Storage and Shipping Temperature: -40°C to 70°C.

NIBP

  • Zoll R Series Patients Type: Adult, Pediatric
  • Zoll R Series Method: Oscillometric
  • Zoll R Series Control: Automatic and manual non-invasive blood pressure measurement.

WiFi Capable

  • WiFi 802.11 a/b/g/n Ambicom-specific 1100C-CF Card P/N 8005-000101-01 compatibility
  • Typical Readiness File Data Size: 750K
  • Zoll Code Data File: 1.2 MB.

Accessories

  • All equipment accessories can be found on the Soma medical equipment accessories and parts website.
  • 8009-0762-01 – OneStep Leads – AHA
  • 9650-0903-01 – Service Manual
  • 8900-000219-01 – OneStep Pediatric CPR Electrode 1ct
  • 8900-000220-01 – OneStep Pediatric CPR Electrode 8ct
  • 8900-0214-01 – Resuscitation Electrode 8ct
  • 8009-0750 – Pacing Cable
  • 8900-0213-01 – CPR Resuscitation Electrode 8ct
  • 8900-0211-01 – Resuscitation Electrode 8ct
  • 8009-0751-01 – SeeThru CPR Simulator

Banner

Questions about our Defibrillators?

Soma Tech Intl offers a wide range of defibrillators. If you have any questions about any of our defibrillators or need a quick quote, call us and one of our knowledgeable sales representatives will help you.

Looking to sell your defibrillators? Soma ensures your facility gets top dollar for your equipment and also gives the option for trade-in credit.

1-800-GET-SOMA
Contact Us