ERBE Erbotom ICC 300

Quotes in less than 24 hours Download Specifications

ERBE Erbotom ICC 300 Features

The ERBE ICC 300 is a high-frequency surgical unit for cutting and coagulation. It’s the price-conscious solution for medium and major surgical procedures. After all, a smooth routine in the operating room depends on maximum safety, a high degree of comfort in use, and perfect functionality – exactly the features of the ICC 300
. And it also meets the special demands of Minimally Invasive Surgery.

  • CUT CONTROL for automatically perfect cuts. Reproducible cut quality with minimum necrosis
  • HlGH CUT with specialized arc regulation for underwater cuts or in fatty structures
  • SOFT coagulation for precisely controlled coagulation
  • No carbonization (SOFT COAG)
  • Far less adhesion and electrode soiling
  • Bipolar coagulation for even greater safety
  • PPS for intelligent support of initial incisions and cutting
  • AUTO START for the greater facility of bipolar coagulation
  • SPRAY coagulation for better surface effects
  • ARGONPLASIVIA coagulation via an integrated plug-in connection

ERBE Erbotom ICC 300 Specifications

Dimensions

  • Dimensions (HxWxD) cm, (In) (Display) 41 x 15.2 x 36.8 (16.1 x 5.9 x 14.5)
  • Weights, kg (lb) (Display) 10 (22)

Electrical

  • Monopolar (Systems) Cut(blend), coagulate
  • Max. energy output ( at rated ohms) (SYSTEMS) More than 250 Watt 300
  • Max voltage, vp-p (Systems) 1200
  • Max. energy output ( at rated ohms) (SYSTEMS) Up to 150 Watt 120
  • Max voltage, vp-p (Systems) 380 soft, 5,600 spray
  • Bipolar (SYSTEMS) Coagulate
  • Max. energy output ( at rated ohms) (SYSTEMS) Up to 150 Watt 120
  • Max voltage, vp-p (Systems) 380

Power Connection

  • Rated power voltage 240 V / 230 V / 115 V / 110 V / 100 V ± 10 %
  • Rated power frequency 50 / 60 Hz
  • Power current 4.0 A at 230 – 240 V / 8.0 A at 100 – 115 V
  • Power consumption in Standby mode 25 watts <?li>
  • Power consumption at max. HF power 620 watts 920 VA
  • Current consumption in Standby mode 150 mA at 230 – 240 V / 300 mA at 100 – 115 V
  • Potential equalization terminal yes
  • Power fuses 2 slow burn, 4 A at 230 – 240 V / 8 A at 100 – 115 V

Environmental Conditions for Storage

  • Temperature -40°C to + 70°C
  • Air humidity, relative 30% to 95%

Environmental Conditions for Operating

  • Temperature +10°C to + 40°C
  • Air humidity, relative 30% to 75%, non-condensing

Cutting, monopolar with automatic voltage control (Auto Cut)

  • HF voltage waveform unmodulated sinusoidal alternating voltage
  • Crest factor C, at RL = 500 ohms C = 1.4 for all settings
  • Rated frequency 330 kHz
  • Maximum HF voltage at load resistance RL = 8 650 VP
  • Dynamic internal impedance in rated load range 0 ohm
  • The constancy of the HF voltage at RL> 350 ohms automatically controlled
  • Cutting quality 4 coagulation effects selectable by key
  • The constancy of the 4 coagulation effects automatically controlled
  • HF rated power 300 watts at RL = 500 ohms
  • HF power limitation 1 to 300 watts in 1 watt steps
  • Setting of the HF power limitation via up/down keys
  • Display of the HF power limitation 7-segment display, 3 decimal places
  • Precision of the HF power limitation +/- 1 digit or +/- 15%
  • Activation of Cutting mode by key or pedal
  • HF connecting sockets 2, electronically locked from one another

Cutting, monopoly with automatic arc control (High Cut)

  • HF voltage waveform unmodulated sinusoidal alternating voltage
  • Crest factor C, at RL = 500 ohms C = 1.4 for all settings
  • The rated frequency of the HF voltage is 330 kHz
  • Maximum HF voltage at load resistance RL = 8 570 VP
  • The constancy of the intensity of the electric arc automatically controlled
  • Cutting quality 4 coagulation effects selectable by key
  • The constancy of the 4 coagulation effects automatically controlled
  • HF rated power 300 watts at RL = 500 ohms
  • HF power limitation 1 to 300 watts in 1 watt steps Setting of the
  • HF power limitation via up/down keys Display of the
  • HF power limitation 7-segment display, 3 decimal places
  • Precision of the HF power limitation +/- 1 digit or +/- 15%
  • PPS (Power Peak System) yes
  • Activation of Cutting mode by key or pedal
  • HF connecting sockets 2, electronically locked from one another

Soft Coagulation (Soft Coag)

  • HF voltage waveform unmodulated sinusoidal alternating voltage
  • Crest factor C, at RL = 500 ohms C = 1.4 for all settings
  • The rated frequency of the HF voltage is 330 kHz
  • The peak value of the HF voltage max. 1900 VP
  • The constancy of the HF voltage automatically controlled
  • HF-rated power 120 watts at 125 ohms
  • HF power limitation (PHFmax.) from 1 watt to 120 watts in 1 watt steps
  • Setting of the HF power limitation via up/down keys
  • Display of the HF power limitation 7-segment display, 3 decimal places
  • Precision of the HF power limitation +/- 1 digit or +/- 15%
  • Activation of Cutting mode by key or pedal
  • HF connecting sockets 2, electronically locked from one another

 

Additiona Features of the ERBE ICC 300 Electrosurgical Unit

Various Coagulation Modes

  • Auto Cut – Automatic open and closed-loop control systems which control and regulate the parameters relevant to the cutting quality so that each respectively selected cutting quality is guaranteed to be reproducible and constant.
  • Endo Cut – For use with snare wires and sphincterotomes.
  • Forced Coagulation – Advantageous if efficient hemostasis is to be achieved with relatively small-surface electrodes, such as TUR resection loops.
  • Soft Coagulation – Can be activated by key or pedal.
  • Bipolar Coagulation – Automatically and constantly controlled. Peak value always remains lower than 200 Vp.

Expandable for Argon Use

  • For units equipped with the Argon Coagulation operating mode, the Erbotom ICC 300 supplies a pulsating high-frequency voltage with peak values up to 4,000 V for Argon Gas coagulation.

Refurbished ERBE Erbotom ICC 300 Electrosurgical Unit

Soma Tech Intl offers Refurbished ERBE Erbotom ICC 300 Electrosurgical unit System up to 50% below OEM prices with the same service and warranty as new. The ERBE Erbotom ICC 300 Electrosurgical Unit is not only technically refurbished but also cosmetically refurbished to make it work and look like new. When the refurbishing process starts on the
ERBE Erboton ICC 300 system is carefully inspected and tested by our own in-house biomedical engineers that are highly skilled and certified to work on ERBE Erbotom ICC 300 Electrosurgical Unit
. If necessary, parts are replaced to make sure that all aspects of the pump are functioning properly. Once everything is in working order, the unit is calibrated back to the original engineering manufacturer specifications. This ensures that the unit is working the same way that it originally left the manufacturer. After the ERBE
system is working like new it undergoes a special cosmetic restoration process. The ERBE Erbotom ICC 300 Electrosurgical Unit is cleaned, minor scratches and dents are repaired, it’s painted, and new decals are replaced if necessary. After this strenuous refurbishing process, the infant warmer functions and looks like new. Soma Tech Intl is ISO 13485 certified and ensures that the best quality is given to all capital medical equipment that they sell. Before the ERBE Erbotom ICC 300
is labeled patient-ready and packed up for shipment, it goes through one last quality control test that makes sure all aspects are again working as they should. If you are looking to buy refurbished ERBE System
contact soma at [email protected] or call us at 1-800-GET-SOMA.

Soma Tech Intl is also actively looking to purchase an ERBE Erbotom ICC 300 Electrosurgical Unit System
If you or your facility is looking to sell your ERBE Erbotom ICC 300, be sure to contact Soma Tech Intl. Soma has a dedicated and experienced purchasing department that helps you through the entire process. Selling your used or pre-owned ERBE Erbutom ICC 300 Electrosurgical Unit System to Soma Tech takes out the middle man in the equation. This also ensures that your medical facility gets top dollar for your equipment. Soma Tech is 100% HIPPA compliant and makes sure that your equipment gets to our facility in the same condition that it left yours. Soma Techn Intl also gives the option for trade-in credit on your medical equipment. If you or your facility is looking to regain floor space for new equipment, or looking to upgrade current equipment, contact Soma Tech’s purchasing department at [email protected] or send us a message on our </a href=”https://www.somatechnology.com/Sell-to-Soma/” target=”_blank”>Sell to Soma website.

Questions about our Electrosurgical Units?

Soma Tech Intl offers a wide range of electrosurgical units. If you have any questions about any of our electrosurgical units, electrosurgical generators, or need a quick quote, call 1-800-GET-SOMA and one of our
knowledgeable sales representatives
will help you.

Banner

Questions about our Electrosurgical Units?

Soma Tech Intl offers a wide range of electrosurgical units. If you have any questions about any of our electrosurgical units or need a quick quote, call us and one of our knowledgeable sales representatives will help you.

Looking to sell your electrosurgical units? Soma ensures your facility gets top dollar for your equipment and also gives the option for trade-in credit.

1-800-GET-SOMA
Contact Us