Philips Avalon FM40 Features

The Avalon FM series fetal and maternal monitors are the first and only to offer automated coincidence detection (cross-channel verification) using Smart Pulse. This innovative feature allows automated maternal pulse detection via the Toco MP transducer without the need to monitor maternal SpO2 or ECG separately. Built-in NST Trace Interpretation, clinical decision support (CDS) application for antepartum monitoring can help to improve efficiency by sorting out the traces that require closer analysis
by the obstetrician. With the Avalon FM series, you have the confidence of continuous data with a backup memory, battery operation, and LAN interface, combined with smart transducers and a color touchscreen.

Cross-channel verification continuously compares all fetal and maternal heart rates to ensure
that each is measured separately, enhancing confidence.

  • Automated coincidence detection with Smart Pulse
  • Optional external touch display, allowing flexible, ergonomic installations
  • Only the Avalon FM series can monitor triplets on a single monitor using the same Ultrasound frequency

Philips Avalon FM40 Specifications

Dimensions

  • Height: 6.8 in (172 mm)
  • Width: 16.5 in (420 mm)
  • Depth: 14.6 in (370 mm)
  • Weight: 19.8 lbs ( <9.0 kg)

Display

  • Monitor Screen: 6.5 in / 16.51 cm
  • Touchscreen Operation: Optional

Battery

  • Operating Time: up to 4 hours

Monitor Operating Environmental Conditions

  • Temperature Range: 0°C to 45°C (32°F to 113°F)
  • Humidity Rage: <95% relative humidity @ 40°C/104°F
  • Altitude Range:-500 to 3000 m/-1640 to 9840 ft.

Monitor Storage Environmental Conditions

  • Temperature Range: -20°C to 60°C (-4°F to 140°F)
  • Humidity Range: <90% relative humidity @ (60°C/140°F)
  • Altitude Range: -500 to 13100 m/-1640 to 43000 ft.

Transducer Operating Environmental Conditions

  • Temperature Rage: 0°C to 40°C (32°F to 104°F)
  • Humidity Range: <95% relative humidity @ 40°C/104°F
  • Altitude Range: -500 to 3000 m/-1640 to 9840 ft.

Transducer Storage Environmental Contditions

  • Temperature Range: -20°C to 60°C (-4°F to 140°F)
  • Humidity Range: <90% relative humidity @ 60°C/140°F
  • Altitude Range: -500 to 13100 m/-1640 to 43000 ft.

SpO2 Environmental Conditions

  • Operating Temperature Range 0°C to 37°C (32°F to 98.6°F)

Power

  • Supply Voltages: 100 VAC to 240 VAC ± 10%
  • Supply Frequency Range: 50 Hz/60 Hz
  • Power consumption (current): 1.3 – 0.7 A

Fetal / Maternal Ultrasound

  • Measurement Method: Ultrasound Pulsed Doppler
  • Measurement Range: US 50 to 240 bpm
  • Display Resolution: 1 bpm
  • Printer Resolution: 1/4 bpm
  • Jitter @ 200 bpm: = 3 bpm
  • Display Update Rate: 1/ Second
  • Average output power P = (4.3 ± 0.4) mW
  • Peak-negative acoustic pressure p_ = (33.9 ± 3.6) kPa
  • Output beam intensity (Iob) (= spatial average – temporal average intensity) Isata = (2.38 ± 0.75) mW/cm<?li>
  • Spatial-peak temporal average intensity I spta = (10.3 ± 2.2) mW/cm2
  • Effective radiating area @ -6 dB 1.81 cm2
  • Beat to Beat Change (max.) for Ultrasound 28 bpm
  • Frequency: 1 MHz ± 100 Hz
  • Signal Range: 3.5 µVpp to 350 µVpp @ 200 Hz

ECG

  • DECG: Single Lead ECG (derived from Fetal Scalp Electrode)
  • MECG: Single Lead ECG (derived from RA and LA electrodes)
  • Measurement Range: 30 to 240 bpm
  • Display Resolution: 1 bpm
  • Recorder Resolution: 1/4 bpm
  • Accuracy: ± 1 bpm or 1%, whichever is greater
  • Beat to Beat Change (max.) 28 bpm
  • Differential Input Impedance > 15MO
  • Electrode Offset Potential Tolerance ± 400 mV
  • Filter Bandwidth 0.8 to 80 Hz
  • Inop Auxiliary Current (Leads Off Detection) < 100 µA
  • Input Signal Range DECG: 20 µVpp to 6 mVpp
  • 20 µVpp to 6 mVpp MECG: 150 µVpp to 6 mVpp
  • Dielectric Strength 1500 Vrms
  • Defibrillator Protection None
  • ESU Protection None

Non-invasive blood pressure

  • Measurement Ranges:
    • Systolic 30 to 270 mmHg (4 to 36 kPa)
    • Diastolic 10 to 245 mmHg (1.5 to 32 kPa)
    • Mean 20 to 255 mmHg (2.5 to 34 kPa)
  • Accuracy:
    • Max. Std. Deviation: 8 mmHg (1.1 kPa)
    • Max. Mean Error: ±5 mmHg (±0.7 kPa)
  • Pulse Rate Range Accuracy:
    • 40 to 300 bpm
    • 40 to 100 bpm: ±5 bpm
    • 101 to 200 bpm: ±5% of reading
    • 201 to 300 bpm: ±10% of reading
  • Measurement Time:
    • Typical at HR > 60bpm
    • Auto/manual: 30 seconds (adult)
    • Maximum time: 180 seconds (adult)
  • Cuff Inflation Time Typical for normal adult cuff: Less than 10 seconds
  • Initial Cuff Inflation Pressure 165 ±15 mmHg
  • Auto Mode Repetition Times 1, 2, 2.5, 3, 5, 10, 15, 20, 30, 45, 60 or 120 minutes

 

Banner