CSZ Blanketrol II

Quotes in less than 24 hours Download Specifications

CSZ Blanketrol II Features

The CSZ Blanketrol II is a Hyper-Hypothermia water therapy system that can be used for the entire patient’s body. The Blanketrol II runs off a Microprocessor that comes with three operational modes, manual mode, auto control mode, and monitor mode. These provide warming and cooling options with high flow rates, allowing for quick regulation of the patient’s core body temperature. Adding water that is lower than the patient’s body temperature to the unit will reduce the patient’s body temperature while adding warm water will increase the patient’s body temperature. The compact size of the system makes it easy to maneuver and set up for any patient at your facility. Another Version of the Blanketrol, is the CSZ Blanketrol III.

  • The unit’s compact size and built-in handle provide ease of use with swivel casters that provide mobility and prevent the unit from tipping. A built-in storage drawer is included in the unit for storage of instruction manuals, probes, connector cables, and hoses.
  • The high water flow rate results in quicker heating or cooling for effective temperature therapy.
  • Three independent safety limits combined with audible and visual alarms to protect your patient.
  • Includes three blanket connections for user convenience.
  • Three operation modes and swivel casters offer ease of use. Operation modes include automatic to control patient temperature, manual to control water temperature, and monitor mode to simply monitor the patient’s temperature.
  • The Blanketrol II also includes a pre-warming/cooling feature to simplify setup.

CSZ Blanketrol II Specifications

Dimensions

  • Height: 36 in
  • Width: 17 in
  • Depth: 17 in
  • Weight: 131 lbs. (59.5 kg.)

Physical Specifications

  • Fluid Temperature Range – Manual: 4°C – 42°C (39.2°F – 107.6°F)
  • Patient Temperature Set Point Range: 30°C to 43.5°C (86°F to 110°F),/li>
  • Refrigerant: Non-CFC (R-134A)
  • Flow Rate: 32 gph (121.1 lph) with the blanket, approx.
  • Heater: 800 Watts
  • Internal By-Pass: Pre-Heating or Pre-Cooling

Thermal System

  • Compressor: 1/3 Hp
  • Heater: 800 Watts
  • Note: Water Cooling rate is Approximately 8°F (4°C) per minute and the heating rate is approximately 6°F (3°C) per minute.

Circulating System

  • Dual Compartment Reservoirs, 2 gallons ( 7.5 liters) total capacity. Error-proof, quick-disconnect coupling. All Circulating components are non-corrosive.

Electrical System

  • Electrical Characteristics
    • 115 60Hz, 10.2 Amps
    • (opt.) 100V, 50/60Hz, 11.6 Amps
    • (opt>) 230 V, 50HZ< 5.2 Amps
    • (opt.) 240 V, 50Hz, 5 Amps
  • Power Cord: 14/3 SJt (100/115 V), Hospital grade plug
  • Leakage Current: Under 300 ua 100/15, Under 500 ua 230/240
  • Circuit Breaker: in Power Switch

Safety System

  • Maximum High Control Setting: 107.6°F (42°C)
  • Primary Safety: 112 °F ± 4°F (44.6 °C ± 2°)
  • Secondary Safety: 115°F ± 4°F (46 °C ± 2°)
  • Maximum Low Control Setting: 39.2 4°F (44.6 °C ± 2°)
  • Primary Safety: 38° F ± 2°F (3°C ± 1°C)
  • Secondary Safety: 34°F ± 2°F (1 °C ± 1)
Banner