Descripción
Covidien ValleyLab FX8 Features
A true next-generation electrosurgical solution, the Valleylab™ FX8 energy platform is powered by our intelligent TissueFect™ tissue-sensing technology. So it reads tissue composition to monitor and adjust to impedance as you work.
The Valleylab™ FX8 energy platform drives electrosurgical handpieces with precision using lower levels of energy.1,2 It’s designed to support a wide range of surgical procedures that require division, hemostasis, and dissection.
Along with cut, coag, and soft coag modes, the Valleylab™ FX8 energy platform enables two unique modes that can bring new possibilities to your procedures:
- Valleylab™ mode provides a unique combination of monopolar hemostasis and division
- Shared coag mode allows two surgeons to work on the same patient at the same time — from a single energy platform
- Precise delivery of monopolar and bipolar energy
- Power control during monopolar energy delivery that’s five times better than the ForceTriad™ energy platform
- Tissue monitoring that’s 2600 times faster than the Force FX™ electrosurgical generator.
- Appropriate thermal spread
Covidien ValleyLab FX8 Electrosurgical Unit Specifications
Dimensions
- Height: 14.6 cm (5.73 in)
- Width: 33.5 cm (13.18in)
- Length: 43.6cm (17.15in)
- Weight: 8.8kg (19.5lb)
General
- Output configuration: Isolated output
- Cooling: Natural convection and fan
- Display: 15.7 cm (6.2 in) LCD touchscreen
- Enclosure: Magnesium
- Mounting
- Valleylab™ universal generator cart (VLFTCRT)
- Operating room boom systems
- Any stable, flat surface such as a table or cart top
Operating Parameters
- Ambient temperature range: 10°C to 40°C (50°F to 104°F)
- Relative humidity: 15% to 85% noncondensing
- Atmospheric pressure: 700 to 1060 millibars
Temperature and Storage
- Ambient temperature range: -10°C to 60°C (14°F to 140°F)
- Relative humidity: 15% to 90% noncondensing
- Atmospheric pressure: 500 to 1060 millibars
Input Power 100-127 VACe
- Line Ranges
- Line Voltage Full Regulation Range: 90–140 VAC
- Line Frequency: 47–63 Hz
- Max VA Nominal Line Voltage: 660 VA
- Max Mains Current: 6.6 ARMS
- Max Heat Dissipation 180 W
- Fuses
- 10 A
- Fuses (2) – 5 mm x 20 mm
- 10 A, 250 V fast blow
Input Power 220-240 VAC
- Line Ranges
- Line Voltage Full Regulation Range: 198-264 VAC
- Line Frequency: 47–63 Hz
- Max VA Nominal Line Voltage: 660 VA
- Max Mains Current: 3.0 ARMS
- Max Heat Dissipation 180 W
- Fuses
- 6.3 A
- Fuses (2) – 5 mm x 20 mm
- 6.3 A, 250 V fast blow
Valoraciones