GE Datex Ohmeda S/5 Anesthesia Monitor
The Datex-Ohmeda S/5 monitor gives you the time and freedom to effectively monitor your patient. It offers everything you need for effortless patient monitoring in the operating room and beyond: a full range of monitoring parameters, intuitive operation, a workstation that can be configured in various ways, seamless device integration, and data transfer.
- Central unit and displays are separate components
- 4 display options with integrated monitor controls: 15″ or 17″ high-resolution color video display on a 10.4″ or 15″ flat panel color LCD screen
- User-configurable screen design and colors
- Supports up to 8 waveforms on the screen
- Support for a second and third user display configurable with different waveforms and numerical information
- Anesthesia-dedicated alarm system
- Alarm display and adjustment page
- 10-minute mini graphical trends referenced to set alarm limits
- Automatic alarm limit functions
- 6 user-configurable modes and special cardiac lead mode
- 24-hour numerical and graphical trends
- Split screen capability to permanently display ST, Spirometry Loop, EEG, EP or MiniTrends
- 4 user-configurable graphical and numerical trends
General Information
- ECG, NIBP, SpO2, Temp, 2-IBP, Rec, CO2 and 5 Agent. color screen
Dimensions
- Height: 13.4 inches (340 mm)
- Width: 13.2 inches (335 mm)
- Depth: 8.1 inches (205 mm)
- Weight (without modules): 24.2 lbs (11 kg)
Voltage
- Nominal voltage and frequency range: 100 to 240v ± 10%, 50/60 Hz
- Maximum energy consumption: 140 VA
- Protection class: Class I
- Grounding: Hospital grade
Monitor
- Screen size: 12.1 diagonal
- Screen type: TFT LCD active matrix color
- Screen resolution: SVGA resolution, 800 x 600
- Number of traces: up to 8
- Screen design and colors: user-configurable. Integrated command card with direct function keys, menu keys, and ComWheel for menu selections and adjustments.
Battery
- Type: Built-in NiMH battery (2 PCS)
- Capacity: Up to 1.5 hours typical on a full charge at 23°C
- Recharge: when connected to the electrical grid
- Charging time: 10 hours at full capacity
- Battery operation alarm
- Indicator: Battery bar displayed on the screen
Tendencies
- Graphic: all parameters, selectable times between 30 min to 72 h
- Numerical: continuously measured parameters sampled every 5 minutes, NIBP and PCWP values updated after each measurement. Additionally, the marker activates the numerical trend update.
- Trend cursor: in graphical and numerical trends