The Capnostream™ 20p patient monitor is built on a legacy of proven performance. For nearly two decades, clinicians have relied on Microstream™ technology for accurate, continuous monitoring of adequate ventilation in intubated and non-intubated patients, from newborns to adults. The new Capnostream™ 20p monitor, with its Apnea-Sat Alert feature, alerts caregivers to repetitive patterns of apnea and oxygen desaturation, revealing a more complete picture of the patient’s respiratory status.
- It supports current standards of care by providing EtCO2 and SpO2 measurements.
- The Capnostream™ 20p patient monitor with Microstream™ includes integrated Smart Alarm Management™ algorithms, a set of proven algorithms to reduce alarms and simplify the use of patient monitoring to improve patient safety and enhance clinical efficiency.
- It protects against low-amplitude «non-breathing» etCO2 excursions, such as talking, crying, or snoring.
- Reduces clinically insignificant alarms.
- It alerts physicians about clinically significant changes in pulse oximetry.
- The IPI algorithm presents a value that demonstrates the respiratory status in real time based on etCO2, RR, PR, and SpO2.
- It records and reports apneas per hour and oxygen desaturation index, and provides the data in trend reports.
- Compatible with central monitoring systems and interfaces with nurse call systems and electronic medical records.
Dimensions
- Unit dimensions: 167mm (h) x 220mm (w) x 192mm (d); (6.6 in (H) x 8.7 in (W) x 7.6 in (D))
- Unit weight: 3.6 kg. 7.94 lb
- Display: 12.1″/30.7 cm diagonal LCD; (NTSC or PAL) with a chemically etched glass layer
- Packaged dimensions: 315mm (H) x 340mm (W) x 285mm (D); 12.4 inches (H) x 13.4 inches (W) x 11.2 inches (D)
Voltage
- Power supply input voltage 100-240 V, 50/60 Hz, ± 10%.
Microstream Capnography
- Units: mmHg, kPa or Vol%.
- Measurement range: etCO2, FiCO2 0-150 mmHg.
- Respiratory rate: 0 bpm-150 bpm.
- Resolution: 1 mmHg.
- Accuracy (± SD): etCO2, FiCO2 0-38 mmHg ± 2 mmHg. 39-150 mmHg ± (5% of reading + 0.08% for each 1 mmHg above 38 mmHg).
- Respiratory rate: 0-70 bpm ± 1 bpm, 71-120 bpm ± 2 bpm, 121-150 bpm ± 3 bpm.
- Flow rate: 50 ml/minute, flow measured by volume.
- Waveform sampling: 50 millisecond intervals.
- Response time: 2.95 seconds (typical).
- Initialization time: 40 seconds (typical).
- Calibration interval: Initial calibration after 1,200 hours of operation, then annually or after 4,000 hours of operation, whichever comes first.
Nellcor Pulse Oximetry
- Measurement range 0-100%.
- Accuracy
- Adult mode: SpO2 range 70%-100% ± 2 digits; SpO2 range 0-69% not specified.
- Neonatal mode: SpO2 range 70%-100% ± 3 digits; SpO2 range 0-69% not specified.
- Pulse rate: Range: 20-250 BPM; Accuracy: ±3 BPM; Sat Sec range: 10, 25, 50, 100.