GE B125 Features

The GE B125 is one of the newest patient monitors from GE. The B125 monitor is known for its simplistic design, with its large 12-inch, captive touch screen. Making it easy for caregivers to interpret and modify patient data promptly. The large numerical or continuous waveforms and easy to see from any angle. The monitor comes standard with advanced parameter monitoring includes; ST-segment analysis, Full arrhythmia analysis, TruSignal SpO2, SuperSTAT NIBP, IBP, Respiratory rate, 3 or 5 Lead ECG, and EtCO2. The GE B125 can be a stand-alone monitor or it can be used along with other devices in a network.

  • ST-Segment and full Arrhythmia analysis, SpO2, NIBP, IBP, RR, ECG, and EtCO2.
  • Records and stores up to 168 hours of monitoring activities across all parameters.
  • Analyses up to 16 types of arrhythmia including AFib.
  • Three-hour battery life for optimal uninterrupted monitoring.
  • A large numeric mode that enables critical parameter visibility even up to 13-feet away.
  • 12.1″ Display

Ge B125 Specifications

Dimensions

  • Height: 11 in (280 mm)
  • Width: 12.5 in (317 mm)
  • Depth: 5.9 in (150 mm)
  • Weight: 9.5 lbs (4.3kg) (with battery)

Display

  • Size: 12.1 in (diagonal)
  • Resolution: 1280 x 800 pixels
  • Number of waveforms: Up to 6
  • Display layout and Color: User-configurable
  • Controls: Capactive Touch Screen & trim knob control and hand keys

ECG

  • Leads available:
    • 3 lead configuration: I, II, III
    • 5 Lead configuraton: I, II, III, aVR, aVf and V
  • Sweep speed: 12.5, 25 or 50 mm/s
  • Gain range: 0.5x, 1x, 2x and 4x
  • Heart rate accuracy 30 to 300 bpm, ± 5% or ±5 bpm, whichever is greater
  • Bandwith: 50/60 Hz power supply
    • Monitor: 0.5 to 40 Hz
    • ST: 0.05 to 40 Hz
    • Diagnostic: 0.05 to 145 Hz
  • Pacemaker detection
    • Range: 2 to 700 mV
    • Pulse width: 0.5 to 2 ms
  • ST-Aaegment Analysis
    • Numeric Range:-9 to +9 mm (-0.9 to +0.9 mV)
    • Accuracy: ± 0.2mm or ±10%, whichever is greater
    • ST Trend: up to 168 h
  • Impedance respiration: range adult – pediatric: 4 to 120 respiratory/min
  • Impedance respiration: range neonate: 4 to 180 resp/min

SpO2

  • GE Ohmeda SpO2
    • Measurement range, Pulse Oximetry: 1 to 100%
    • Pulse rate: 30 to 250 bpm
    • Measurement accuracy – saturation
      • Without Motion-Adult/Pediatric Finger sensor: 70 to 100% ± 2%
      • Without Motion-Neonate: 70 to 100% ± 3%
      • With Motion-Adult/Pediatric/Neonate: 70 to 100% ± 3%
      • Low Perfusion-Adult/Pediatric: 70 to 100% ±3% (1~69% unspecified)
    • Pulse rate: without motion: ± 2 bpm ( Adult/ Pediatric/ Neonate)
  • Nellcor Oximax
    • Measurement Range – Pulse Oximetry: 1 to 100%
    • Pulse Rate: 20 to 250 bpm
    • Measurement accuracy – saturation:
      • Adult: 70 to 100% ±2%
      • Neonatal: 70 to 100% ±3%
      • Low Perfusion: 70 to 100% ±2% (0~69% unspecified)
    • Pulse rate: 20 to 250 bpm
  • Masimo
    • Measurement Range – Pulse Oximetry: 1 to 100%
    • Pulse Rate: 20 to 250 bpm
    • Measurement accuracy – Saturation
      • Without Motion-Adult/Pediatric: 70 to 100% ±2%
      • Without Motion-Neonate: 70 to 100% ±3%
      • With Motion-Adult/Pediatric/Neonate: 70 to 100% ±3%
      • Low Perfusion: 70 to 100% ±2%(0~69% unspecified)
    • Pulse Rate without motion: ±3 bpm
    • Pulse rate with motion: ±5bpm

NIBP

  • Measurement Technique: Oscillometric with step deflation
  • Modes: Manual, automatic, and slat
  • NIBP Measurement Range:
    • Systolic Adult / Pediatric: 30 to 290 mmHg
    • Systolic Neonates: 3 to 140 mmHg
    • MAP Adult / Pediatric: 20 to 250 mmHg
    • Diastolic adult / pediatric: 1o to 22 mmHg
    • Diastolic neonate: 10 to 11o mmHg
    • Defat initial inflation pressure
      • Adult / Pediatric: 135 ±15 mmHg
      • Neonate: 100 ±15 mmHg
    • Maximum Determination Time
      • Adult/ Pediatric : 2m
      • Neonate: 85 s
    • Over Pressure Monitor
      • Adult /Pediatric: 294 ±6 to 330 mmHg
      • Neonate: 147 ±3 to 165 mmHg

Paper Recorder

  • Method: Thermal dot array
  • Horizontal Resolutions: 24 dots/mm (600 dpi)
  • Vertical Resolution: 8 dots/mm (200 dpi)
  • Waveforms Selectable: 1, 2, or 3 waveforms
  • Numerics: HR, SpO2, NIBP, IBP1, IBP2, EtCO2/FiCO2, T1, T2, Resp
  • Graphical Trend Printout: HR, ST, IBP1, IBP2, NIBP, SpO2, Pleth, CO2, Resp, T1+T2
  • Paper width: 50 mm, printing width 48 mm
  • Paper Speed: 6.25, 12.5, 25 mm/s

Operating Conditions

  • Temperature: 5 to 40°C (41 to 104°F)
  • Relative Humidity: 20 to 90% noncondensing
  • Atmospheric Pressure: 700 to 1060 hPa (525 to 795 mmHg)

Storage and Transport Conditions

  • Temperature: -20 to 60°C (-4 to 140°F)
  • Relative Humidity: 10 to 90% noncondensing
  • Atmospheric Pressure: 700 to 1060 hPa (525 to 795 mmHg)

Power Specifications

  • AC Input: 100 to 240V ±10%, 50/60 Hz, 150VA
  • Protection: Class I
  • Battery: Exchangeable lithium-ion, 1 pcs max
  • Charging Time: < 4 h to 90% capacity
  • Run Time: > 3 h
Banner

¿Pregunte sobre nuestras Monitor multiparametro para UTI?

Soma Tech Intl ofrece una amplia gama de Monitor multiparametro para UTI. Si tiene alguna pregunta sobre cualquiera de nuestros equipos medicos o necesita una cotización rápida, llámenos y uno de nuestros expertos representantes de ventas lo ayudará.

¿Quiere vender sus Monitor multiparametro para UTI? Soma le garantiza que usted obtendrá el mejor precio por su equipo y también ofrece la opción de crédito de intercambio.

1-800-GET-SOMA
Contact Us