Defibtech DDU-100 Lifeline

Cotizaciones en menos de 24 horas

Defibtech DDU-100 Lifeline AED Features

The Defibtech DDU-100 Lifeline is a semi-automatic external defibrillator (AED). The Defibtech DDU-100 is a very simple straightforward AED. The Defibrillator system offers voice prompts and visual indicators to make it easy to operate. The Defibtech AED has two user control, on/off and Shock. The Shock button is only enabled when a shockable rhythm is detected. The System can deliver 150 joules of energy to an adult patient when a shock is required, and 50 joules of energy to child and infant patients. The Defibtech DDU-100 Lifeline AED runs off a replaceable, non-rechargeable lithium battery.

  • Automatically analyzes the patient’s ECG.
  • Determines whether a shockable rhythm is present.
  • Prompts the operator to press the Shock button when the device is ready and recommended.
  • Defibrillation energy is delivered as an impedance compensated biphasic truncated exponential waveform.
  • Delivers 150J of defibrillation energy through a patient impedance range of 25-180 ohms or 50J of defibrillation energy when using the child/infant pads.
  • Replaceable non-rechargeable lithium battery pack.

Defibtech Lifeline DDU-100 Specifications

Dimensions

  • Height: 11.8” (30 cm)
  • Width: 8.5” (22 cm)
  • Depth: 2.7” (7 cm)
  • Weight: 4.4 lbs (2 kg)

General

  • Waveform: Biphasic Truncated Exponential
  • Energy: Adult: 150 Joules; Infant/child: 50 Joules
  • Charge Control: Automatic by Patient Analysis System
  • Charge Time From Shock-Advised: Typically < 6 seconds with a fresh DBP 2800 battery pack and < 9 seconds with a fresh DBP-1400 battery pack. Charge time may increase at the end of battery life and for temperatures below 10°C.
  • Charge Complete Indication: SHOCK button flashing; “Press Shock button” voice prompt
  • Shock Delivery: Shock is delivered by a single SHOCK button

Disarm Automatic

  • If Patient Analysis System decides rhythm is no longer shockable, or Within 30 seconds after Charge complete, if an operator has not pressed the SHOCK button, or
  • If defibrillation pads are removed from the patient or unplugged from the unit

Disarm Manual

  • If the operator presses the OFF/DISARM button at any time to disarm and turn off the device.

High-Capacity Battery

  • Model Number: DBP-2800
  • Main Battery Type: 15VDC, 2800 mAh, Lithium/Manganese Dioxide. Disposable, recyclable, non-rechargeable
  • Capacity: 300 shocks or 16 hours of continuous operation
  • Charge Time: < 6 seconds
  • Standby-life (installed in the unit): 7 years
  • Active Status Indicator (ASI) Battery: 9VDC, 1200 mAh, Lithium Manganese Dioxide. Disposable, recyclable, non-rechargeable.
  • ASI Battery Standby-Life (after installation): >1 year

Standard Battery

  • Model Number: DBP-1400
  • Main Battery Type: 15VDC, 1400 mAh, Lithium/Manganese Dioxide. Disposable, recyclable, non-rechargeable.
  • Capacity: 125 shocks or 8 hours of continuous operation.
  • Charge Time: < 9 seconds
  • Standby-life (installed in the unit): 5 years
  • Active Status Indicator (ASI) Battery: 9VDC, 1200 mAh, Lithium/ Manganese Dioxide. Disposable, recyclable, non-rechargeable.
  • ASI Battery Standby-Life (after installation): >1 year