GE MAC 5 Features
The GE MAC 5 EKG machine is a portable electrocardiograph that can capture 3, 6, or 12-lead electrocardiograms on adult and pediatric patients. It features interpretive analysis and the ability to connect to a network using a wired LAN connection or wireless WiFi access point. The MAC 5 comes in 3 different models: the Mac 5 A4 with an A4 printer, the MAC 5 A5 with an A5 printer, and the MAC 5 Lite with no printer.
- 3, 6, and 12-lead ECG
- 8.9-inch touchscreen
- Gender and pediatric-specific interpretations
- Easy-to-clean design
GE MAC 5 Specifications
GE MAC 5 A4 Dimensions
- Height: 12.4 in (31.5 cm)
- Width: 14.2 in (36 cm)
- Depth: 4.5 in (11.4 cm)
- Weight: 7.9 lb (3.6 kg)
GE MAC 5 A5 Dimensions
- Height: 12.4 in (31.5 cm)
- Width: 10.2 in (26 cm)
- Depth: 4.3 in (10.8 cm)
- Weight: 6.6 lb (3 kg)
GE MAC 5 Lite Dimensions
- Height: 12.2 in (30.9 cm)
- Width: 10.2 in (26 cm)
- Depth: 3.4 in (8.4 cm)
- Weight: 4.4 lb (2 kg)
Display
- Display: 8.9 in (22.6 cm) diagonal graphics LED backlit full HD
- Touchscreen: Projected Capacitative (PCAP) multipoint touch input that works while wearing medical exam gloves
- Resolution: 892 x 558 pixels, with waveform enhancement
Printer (A4 / A5 Models)
- Technology: Integrated thermal dot array
- Writer Speed: 5, 12.5, 25, and 55 mm/s
- Number of Traces: 3, 6, 12
- Paper Type: Thermal Z-fold
- Paper Capacity: Up to 150 sheets
Battery
- Type: Replaceable and rechargeable internal battery
- Charge Time: Approx. 240 minutes
- Capacity: Shall be able to acquire and print a single page 12-lead resting ECG report every 15 minutes for a minimum of 3 hours
External USB Barcode Reader (Optional)
- Symbologies: Code-128, PDF417, Code 39, Interleaved Code 2 of 5, and Data Matrix symbology for characters A-Z and 0-9 for all supported languages
Electrical
- Power Supply: AC mains or battery operation
- Input Voltage: 100-240 VAC±10%
- Input Current Range: A @100V to 625mA @240V
- Input Frequency: 50/60 Hz ±3 Hz
ECG Data Acquisition
- Signal Input: Type: CF, Defibrillation-Proof; Defibrillation Protection: Per IE C 60601-2-25:2011
- Dynamic Range: AC Differential ± 10mV, DC offset ±600 mV
- Common Mode Rejection: >125 dB (> 100dB with AC filter disabled)
- Input Impedance: >50 MO @ 10 Hz, defibrillator protected
- Patient Leakage Current: <10 µA
- Detection of Pacemaker Pulse: Duration: 0.2 ms to 2.1 ms; Amplitude: 2 mV to 700 mV; Separation: 1 ms or greater
- Page Annotation: Dedicated pace channel on display and printed reports
- Pace Digital Sampling Rate: 75,000 samples/second per channel
ECG Data Processing
- ECG Interpretation: Marquette 12SL ECG Analysis Program for Adults and Pediatrics
- Computerized Measurements: 12-lead analysis
- Heart Rate Meter: 30 to 300 beats per minute (BPM) with an accuracy of ±10% or 5 BPM, whichever is greater. Heart rates outside this range will not display
- ECG Data Formats: GE Hi-Fidelity ECG, XML
- Pre-acquisition: Provides 10 seconds of instantaneous ECG acquisition
- Digital Rhythm: Up to 5 minutes of continuous rhythm storage (exportable as a PDF)
- Storage: 300 records consisting of 10-second Resting ECG records and Digital Rhythm records on the device’s internal memory
- External Storage: USB-compliant flash drive supporting the FAT32 file system
- Downsampled ECG Waveform: Bandwidth: 0.04 to 300 Hz; Sample rate: 2 kbps;
- Resolution: 1.22 µV
- Analyzed ECG Waveform: Bandwidth: 0.04, 0.56 ZPD to 300 Hz; Sample Rate: 2 kbps; Resolution: 1.22 µV
- Additional Report Filters: 20 Hz, 40Hz, 100 Hz, 150 Hz, 300Hz
- Channels: Up to 12 channels, skew between channels: <100 µS