Welch Allyn PIC 50

Quotes in less than 24 hours Download Specifications

Welch Allyn PIC 50 Defibrillator Features

The Welch Allyn PIC 50 is a defibrillator that offers monitoring features. The Welch Allyn PIC 50 offers Biphasic waveforms, 12 lead ECG, Pacing, oxygen levels, invasive and non-invasive blood pressure, end-tidal CO2, and patient temperature. The PIC50 is a portable semi-automatic defibrillator that can be used for both hospital and in-transit use. The Welch Allyn defibrillator was designed for easy use with a 6.5-inch color display screen, a built-in ECG printer, and a clearly defined button. The System comes with defibrillator paddles for adult patients.

  • Portable lightweight design that weighs 10lbs
  • Rugged design makes it ideal for both in and out-of-hospital use
  • Easy-to-use system with custom configurations
  • Communication optional with cellular transition, fax, and data card transfers

Welch Allyn PIC 50 Defibrillator Specifications

Dimensions

  • Height: 5.4in (13.7 cm)
  • Width:12.6 in (32 cm)
  • Length: 13.2 in (33.5 cm)
  • Weight: 13.2 lb (6.0 kg) (with battery)

Defibrillator

  • Waveform: Biphasic: Low-current, impedance-compensated, optimized terminating current, truncated exponential biphasic
  • External Select: External: 2, 5, 7, 10, 20, 30, 50, 70, 100, 150, 200, 300, 360 JM
  • Charge Time: < 10 seconds @ 360 J (typical)
  • Charge indicator: : Audible and graphic
  • Output: Adult paddles, internal paddles, pediatrics adapter, and multipurpose hands-free adapter available.
  • Synchronizer: Delivers energy within 60 msec
  • Disarm: Front panel Switch

Monitor/ Display

  • Input: 12-lead PIC: 12-lead patient cable, paddles, or multipurpose hands-free adapter
  • Size: 6.4″, (16.3 cm) diagonal, non-fade
  • Type: Color active matrix TFT display
  • Display Resolution: 640 x 480 pixels
  • Sweep Speed: 25 mm/sec
  • Lead Selections: Paddles (Pads), I, II, III, AVR, AVL, AVF, V</li.
  • Frequency Response: (User-selectable)
    • 2 to 20 Hz Limited mode
    • 0.67 to 40 Hz Monitor mode
    • 0.25 to 40 Hz Filtered Diagnostic mode
    • 0.05 to 150 Hz Diagnostic mode
  • Common Mode Rejection: 90 dB (Complies with AAMI EC13-1992 section 3.2.9.10)
  • Heart Rate Meter: 20 to 300 BPM
  • Heart Rate Alarms: User-selectable Size: 0.125, 0.25, 0.5, 1, 2, 4 cm/mV and auto-sizing
  • ECG Output: 1 V/mV

Recorder

  • Type: High-resolution thermal array
  • Annotation: Time, date, ECG lead, ECG gain, heart rate, defibrillation and pacing parameters, and treatment summary ACLS events
  • Paper Width: 50 mm
  • Paper Speed: 25 mm/sec, 50 mm/sec
  • Delay: 6 seconds
  • Frequency Response: Automatically set to monitor’s frequency response
  • Treatment Summary: 7 switches to record key ACLS events (IV, INTUB, EPI, LIDO, ATROP, etc.). Automatically logs into memory the type of event, time, and ECG sample
  • Tx Summary Log: 60 ECG events or 4,000 non-ECG events
  • Record Modes: Manual and automatic
  • Voice MemoTM (optional): Quickly and easily records the operator’s voice into the internal memory
  • Record Modes: Manual & automatic (user configurable)

Battery

  • High-Capacity Type: NiMH 12 V Welch Allyn SuperPac
  • Low-Battery Indicator: Flashing low-battery icon on display
  • Multiple Paired Contacts: Ensure quick, error-free insertion and backup reliability
  • Recharge Rate*: 80% in 7.5 hours, 100% in 9.5 hours

12 Lead

  • Acquired 12-lead: I, II, III, AVR, AVL, AVF, V1- V6 acquired simultaneously
  • Transmitting: Optional fax/modem; 10 seconds of ECG (12 leads)
  • Printing: 3-second samples of 12 leads
  • 12-lead Interpretation (optional): Provides interpretation statements and measurements. Also provides Welch Allyn’s exclusive on-screen display and review of interpretation results

Pulse Oximeter

  • Measurement Range: 31 to 99%
  • Heart Rate Range: 30 to 250 BPM ± 5% or 5 BPM (whichever is less)
  • Accuracy: 81 to 99% = ± 2%, 70 to 80% = 3%, 0 to 69% unspecified
  • Update Period: Every pulse after 4 valid pulses
  • Messages: Check Probe, Searching, Signal OK, No Signal, Poor Signal, Alarms Enabled/Disabled (icon), Testing
  • Alarm Limits: 60 lower limits, 70 upper limits to 99%.
  • Probe: Reusable finger or multisite probe

Pacer (option)

  • Type: External transcutaneous pacing. <?li>
  • Pacer Rate: 30 to 180 BPM ± 5%
  • Output Current: 30 to 180 mA ± 10% or 5 mA (whichever is greater)
  • Modes: Demand and asynchronous
  • Status Indicators: ECG lead fault, pace lead fault, pace marker on monitor and chart, start/stop LED
  • Pulse Type: Rectangular, constant current
  • Pulse Width: 20 ms
  • Refractory Period: 250 ms
  • Output Protection: 360 joules

IBP (option)

  • 2 Channels: Systolic, Diastolic, MAP, and waveform displays
  • Pressure Range: -30 to 300 mmHg
  • Accuracy: 2%
  • Transducer: 5µV/V/mmHg
  • Offset Imbalance: ±125 mmHg, including transducer offset

CO2

  • Mainstream CO2 : C02 and EtCO2 display
  • Range: 0 to 76 mmHg
  • Accuracy: + 4 mmHg (< 40 mmHg), + 10% (> 40 mmHg)
  • Resolution: 1 mmHg
  • Respiration Range: 3 – 60 BPM

NIBP (option)

  • Technique: Noninvasive oscillometric method, referenced to auscultatory standard, using the 5th Korotkoff sound as the diastolic reference
  • Operating Modes: Automatic and manual
  • Initial Cuff Inflation: Adult and child: 154 mmHg, Infant: 125 mmHg
  • Automatic Cycle Times: 1, 2, 3, 5, 10, 30-minute intervals
  • Blood Pressure Range: Systolic: 30 to 245 mmHg, Diastolic: 10 to 210 mmHg
  • Heart Rate Range: 30 to 200 BPM ± 10% or 5 BPM
  • Blood Pressure Accuracy: +5 mmHg mean error, +8 mmHg standard deviation
  • Blood Pressure Validation: Blood pressure measurements determined with this device are equivalent to those obtained by a trained observer using the cuff/stethoscope auscultation method within the limits prescribed by the American National Standards Institute
  • Alarms: Artifact, check cuff, tighten cuff, cuff leak, motion
  • Alarm Limits: Systolic: 30 to 250 mmHg Diastolic: 25 to 250 mmHg

Temperature (option with NIBP)

  • Measurement Range: 32° to 122°F (0° to 50°C)
  • Accuracy: 0.3°F (± 0.2°C)
  • Resolution: 0.1°F (0. °C)
  • Scale: Fahrenheit or Celsius
  • Probe: YSI Series 400
  • Alarm Limits: High 99-109 °F (37° to 43°C); low 85-98°F (29° to 36°C) alarm limits

AED (option)

  • Configurable: Basic & Basic+ modes
  • Operation: Continuous analysis of ECG advises shock on VFIB, shockable VTACH <?li>
  • Energy Sequence: User-configurable
    • Monophasic: 200/200/360 J or 200/300/360 J
    • Biphasic: 150/200/300/360 J
  • Factory Default:
    • Monophasic: 200/300/360 J
    • Biphasic: 150/200/300 J
  • Prompts: Audible voice and visual text prompts