Mortara ELI 150C
24-Hour QuotesPricing up to 50% below the cost of new,
with service and warranty included.
Mortara ELI 150C Features
The Mortara ELI 150C is a 12-lead resting electrocardiograph used for ECG acquisition, analysis, and storage in clinical environments. It supports portable operation and network connectivity for transferring ECG data to hospital systems. The unit records, processes, and displays ECG waveforms and supports automated interpretation using built-in analysis software. It can be used with adult and pediatric patients.
- Portable System: Compact ECG unit weighing approximately 7.2 lb with a sealed elastomer keyboard and backlit color display. Waveforms can be viewed in 3, 4+4, or 6+6 lead formats.
- Acquisition Module Options: Compatible with the AM12 acquisition module or the Wireless Acquisition Module (WAM) for cable-free ECG acquisition and rhythm printing.
- ECG Interpretation Software: Includes the VERITAS resting ECG interpretation algorithm, which applies adult- and pediatric-specific criteria for preliminary analysis.
- Signal Processing: Uses Best 10 technology to select a 10-second ECG segment with reduced noise for display and printing.
- Data Security Controls: Supports user authentication, audit trail logging, and 256-bit AES encryption to assist with compliance requirements.
- Network Connectivity: Supports LAN and WLAN communication using HL7, DICOM, PDF, and XML formats for transfer to EHR, PACS, or CVIS systems.
- Recording Resolution: Uses a 20-bit analog-to-digital converter with a sampling rate of up to 40,000 samples per second per channel for waveform recording and pacemaker spike detection.
Mortara ELI 150C Specifications
Dimensions
- Height: 3.75 in (9.5 cm)
- Width: 11.25 in (28.6 cm)
- Depth: 11.5 in (29.2 cm)
- Weight: 7.2 lbs (3.3 kg) including battery (without paper).
Power
- AC Input: Universal AC power supply, 100-240 VAC, 50/60 Hz, 110 VA.
- Battery: Internal rechargeable battery for portable operation.
ECG Acquisition & Performance
- Input Channels: Simultaneous acquisition of all 12 leads.
- Standard Leads: I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6.
- Digital Sampling Rate: 40,000 samples/sec/channel (for pacemaker detection); 1,000 samples/sec/channel (for recording/analysis).
- A/D Conversion: 20 bits (1.17 microvolt LSB).
- Frequency Response: 0.05 to 300 Hz.
- Input Dynamic Range/CMRR/Impedance: Meets or exceeds IEC 60601-2-25 requirements.
Display & Printing
- Display: Backlit, 1/4 VGA color LCD (320 x 240 pixels). Waveform display in 3, 4+4, or 6+6 lead presentations.
- Printer: Computer-controlled dot array thermal printer (8 dots/mm).
- Paper: Perforated double Z-fold thermal paper; 108 mm (4.25″) wide, 200 sheets.
- Print Speeds: 5, 10, 25, or 50 mm/s.
- Gain Settings: 5, 10, or 20 mm/mV.
- Report Formats: Standard or Cabrera; 3, 3+1, 3+3, or 6 channel.
Data Management & Safety
- ECG Storage: Internal storage for up to 40 ECGs; optional expansion to 200 ECGs.
- Connectivity: Bidirectional via LAN or WLAN using XML, PDF, DICOM, and HL7 protocols.
- Safety Standards: Patient and chassis leakage currents meet or exceed ANSI/AAMI/IEC 60601-1 requirements.
Filters
-
Includes high-performance baseline filter, selectable AC interference filter (50/60 Hz), and low-pass filters (40 Hz, 150 Hz, or 300 Hz).
Refurbished Mortara ELI 150C EKG Machine
Soma Tech Intl offers the Mortara ELI 150C EKG Machine up to 50% below OEM prices with the same service and warranty as new. The refurbished Mortara ELI 150C is not only technically refurbished but also cosmetically refurbished to make it work and look like new. When the refurbishing process starts on the ELI 150C, it is carefully inspected and tested by our own in-house biomedical engineers who are highly skilled and certified to work on EKG machines. If necessary, parts are replaced to ensure that all aspects of the ELI 150C meet the original manufacturer’s specifications. This ensures that the unit operates just as it did when it left the manufacturer.
Once the Mortara device is functioning like new, it undergoes a specialized cosmetic restoration. The unit is cleaned, minor scratches and dents are repaired, it’s repainted, and new decals are added if necessary. After this thorough refurbishment, the unit both functions and appears like new. Soma Tech Intl is ISO 13485 certified, ensuring top-quality standards in every piece of medical equipment sold. Before the Mortara ELI 150C is labeled patient-ready and prepared for shipment, it goes through a final quality control check to confirm all aspects are working as they should. If you are interested in a refurbished EKG machine, contact Soma at soma@somatechnology.com or call us at 1-800-438-7662.
Soma Tech Intl actively seeks to purchase EKG machines. If you or your facility are looking to sell your Mortara ELI 150C, contact Soma Tech Intl. Soma’s experienced purchasing team will guide you through the process to ensure your medical facility receives the best value for your equipment. Soma is fully HIPAA-compliant and ensures your equipment arrives in the same condition it left your facility. Soma also offers trade-in credits for medical equipment. If you have equipment to sell, send us a message here.