B Braun Outlook 100ES
24-Hour QuotesPricing up to 50% below the cost of new,
with service and warranty included. Download Specifications
B Braun Outlook 100ES Features
Equipped with a color-coded keypad and audible key clicks, the safe, reliable, and easy to use Outlook® 100ES infusion pump provides comprehensive “at-a-glance” viewing of infusion status on a large color display.
Designed to help prevent IV medication administration errors, the Outlook® 100ES Safety Infusion Pump System’s extensive customizable DoseGuard™ drug library allows you to configure the pump to match your clinical practice.
- Intuitive: A “brain” in each pump helps prevent channel confusion and catastrophic pump failures
- Versatile: Vertical stacking takes less space
- + KeyGuard™: Non-numeric keypad with simple arrows to help prevent manual programming errors related to zero vs. decimal entry
- Air Trapping Cassette: Catches up to 3 mLs of air helping to reduce nuisance air-in-line alarms
- Proven delivery mechanism providing reliable and accurate operation
- Real-time display of dose and limit values during infusion for titration safety
- Active alphabetic matrix for quick and easy drug selection
- Loading Dose and On Demand Bolus configurations for all DoseGuard Drugs
B Braun Outlook 100ES Specifications
Dimensions
- Width: 30.5 cm (12 in)
- Height: 11.4 cm (4.5 in)
- Depth: 22.2 cm (8.75 in)
- Weight: 5 kg (11 lbs)
Power
- Power: 120 VAC, 3A, 60 Hz
- Grounding Resistance: Meets or Exceeds UL Standard 2601-1
- Leakage Current: Meets or Exceeds UL Standard 2601-1
General
- Mechanism: Positive Pressure Displacement Reservoir
- UL Certified: Yes. Meets or exceeds 2601-1, IPX1 drip rating
- Battery: 12V NiMH
- Battery Operating Time: 7 hrs at 25 ml/hr
- Rate Range: 0.1 to 999.9 mL/hr in 0.1 mL increments
- Micro Mode: 0.1 to 99.9 mL/hr in 0.1 mL increments
- Volume to be Delivered Range: 0.1 to 9,999.9 mL in 0.1 mL increments
- Occlusion Pressure: Adjustable Settings: 75, 100, 200, 300, 400, 500, and 750 mmHg
- KVO Rate: 3 mL/hr or selected rate whichever is lower
- Volumetric Accuracy: ± 5%
- Alarms: Air-in-line, Hold Time Exceeded, Low Battery Warning, Plug In Alarm, DoseGuard™ Alert, RateGuard™ Alert, Downstream Occlusion, Check Set, Bag Empty, Upstream Occlusion, System Error
- Tubing: Horizon® Pump Sets
- Modes/Features: DOSE Mode with DoseGuard™ Drug Library, Primary and Piggyback Modes with RateGuard™ Solution Library, Loading Dose and On-demand Bolus Modes, Profile Mode, Program Mode, Set Time and Volume, Schedule Next Run, Micro Mode, q(x) Mode, Epidural Mode
- Custom Configuration Software: Outlook® ES Pump Configuration Editor Software (Required—One-time fee, unlimited use license)