Impact Uni-Vent 754 Features
The Impact Univent 754 is a portable ventilator designed for various PEEP ventilation modes, including Assist Control, SIMV, CPAP, and CMV. It features an integrated compressor and blender, making it completely self-contained and eliminating the need for external gas sources. Suitable for patients from Neonatal to adult, it includes a backlit LCD screen for monitoring vital signs and alarm settings. Powered by an internal rechargeable battery with optional AC power connectivity, the system also offers an interactive demo and teaching mode. Additionally, it is certified for EMI/RFI and Air Medical use.
- Compact and lightweight
- Operating Modes: ACV, SIMV, CPA, and CMV
- Fully self-contained system: integrated compressor and oxygen blender eliminate the need for external gas sources during operation.
- Wide patient range: suitable for neonatal, pediatric, and adult patients across diverse care settings.
- Clear visual monitoring: backlit LCD screen displays vital parameters, settings, and alarm conditions for quick clinical assessment.
- Flexible power options: internal rechargeable battery with optional AC power ensures uninterrupted ventilation during transport.
Impact Uni-Vent 754 Specifications
Operating Modes
- ACV – with/without PEEP, SIGH, and PRESSURE PLATEAU
- SIMV – with/without PEEP SIGH and PRESSURE PLATEAU
- CPAP – with/without PEEP
- CMV – for Apnea backup of ACV, SIMV, and CPAP
General
- Deliverable Gas(es): Air (Internal Compressor, External 50-PSI); Oxygen (External 50-PSI)
- Tidal Volume Range: 0 to 3000 mL
- I:E Ratio: 1:1 Preset, Adjustable 1:1 to 1:599
- Sensitivity: default, 1.5 to 2.0 cmH2O; Adjustable, 1.0 to 6.0 cmH2O
- Flow Rate: adjustable, 0 to approximately 60 LPM (0 to approximately 1000 ml/sec)
- Ventilation Rate: adjustable, 1 to 150 bpm, in 1bpm increments (±1 digit on the display panel)
- Inspiration Time: adjustable, 0.1 to 3 sec, in 0.1-sec increments (±1 digit on the display panel); 1:2 I: E RATION preset
- FiO2: adjustable, 21% to 100% (Internal Air/Oxygen Mixer), resolution in 1% increments, accurate to within ±10%
- Low Pressure Alarm: adjustable, 0 to 50 cmH2O, in 1 cmH2O increments (±1 digit on the display panel)
- High Pressure Alarm: adjustable, 15 to 100 cmH2O, in 1 cmH2O increments (±1 digit on the display panel)
- Peek Inspiratory Pressure Relief: adjustable, 15 to 100 cmH2O, in 1 cmH2O increments (±1 digit on the display panel)
- Pressure Plateau Range: 5 to 90 cmH2O (referenced to HIGH PRESSURE ALARM set-point)
- Assist/SIMV: sensitivity Default, 1.5 to 2.0 cm H2O below end pressure
- PEEP Program: range 1 to 20 cmH2O, in 1 cmH2O increments (±1 digit on the display panel)
- Sigh duration: 150% Vt (truncated to a combined maximum of 3 seconds)
- Liquid Crystal Display: External Air, Sigh, PEEP, Pressure Plateau, High-Pressure Alarm Setting, Low-Pressure Alarm Setting, Ventilation Rate, Inspiration time/I: E Ratio, Tidal Volume, Air/Oxygen Mixer, Mode, Inspiration/Exhalation, Power, Pear Airway Pressure, Mean Airway Pressure, Digital Bar Graph, High/Low, Airway Pressure Alarm Set point Indicators, Paw.
- LED Indicators: Charge/Alarm/System Failure
- LED Digital Bar Graph Range: -10 to 100 cmH2O
- LCD Alarm Display: Battery Low, External Power Low, Low Pressure, O2 Low/Fail, Disconnect, High Pressure, Apnea, Vt, High PEEP, Ext Air Low/Fail, FIO2, Inverse I/E, Comp, Pressure Alarm Settings, Transducer Calibration, Transducer Calibration Abort, System Failure, Ventilator Fail, Inspiration Time Truncated to 3-seconds, Plateau Volume, Preventative Maintenance Due, Vt Settings, Extended Non-Use/ Storage, External Power Failure
Refurbished Impact UniVent 754 Ventilator
Soma Tech Intl offers the Impact Univent 754 ventilator at up to 50% below OEM pricing, while providing the same service and warranty as a new unit. Each Univent 754 is thoroughly refurbished both technically and cosmetically to ensure reliable ventilation performance in transport, emergency, and clinical care environments. During the refurbishment process, the Impact Univent 754 is carefully inspected and tested by our certified in-house biomedical engineers, who are experienced with portable and critical care ventilators. Any components that do not meet the original manufacturer’s specifications, such as pneumatic systems, internal electronics, valves, sensors, alarms, or power components, are repaired or replaced with OEM-quality parts to ensure safe and consistent operation.
Once the technical refurbishment is complete, the Univent 754 ventilator undergoes a comprehensive cosmetic restoration. The unit is thoroughly cleaned, minor scratches and surface wear are repaired, housings are refinished if needed, and new decals or labels are applied when necessary. After this detailed process, the Impact Univent ventilator functions and appears like new. Soma Tech Intl is ISO 13485 certified, ensuring that every refurbished medical device meets the highest standards of quality, safety, and reliability. Before shipment, each unit is labeled patient-ready and undergoes a final quality control inspection to verify proper operation. If you are interested in purchasing a refurbished Impact Univent 754 ventilator, contact Soma at soma@somatechnology.com or call 1-800-438-7662.
Soma Tech Intl is also actively seeking to purchase Impact Univent 754 ventilators. If your facility is upgrading respiratory equipment or has units available for sale, our experienced purchasing department will guide you through the entire process. Selling directly to Soma removes the middleman and helps ensure your organization receives top market value. Soma is 100% HIPAA compliant and guarantees secure handling and transportation of all equipment. Trade-in credit options are also available. To sell your equipment, send us a message here.