Vyaire 3100b Ventilator Features
The Vyaire 3100B is a High Frequency Oscillatory Ventilatory. The Vyaire 3100B is Base on the Same platform as the Vyaire 3100A. The 3100 B added additional features for adult patients. The Vyaire can be used on patients that weigh over 77 lb (35kg). The system is used for patients that suffer from acute respiratory distress. The ventilator provides tools that can protect the lungs from damage while it is continuously expending pressure and volume to inflate the lungs. This helps to prevent the patient from going into respiratory failure.
- Produces an active exhalation, which is essential at high-frequency respiratory rates to prevent air trapping that may occur with passive exhalation.
- Offers patented technology and a highly reliable, electromagnetically driven piston that distinguishes it from other high-frequency ventilators.
- Permits variable I:E ratios, which are desirable for managing ventilation.
Vyaire 3100b Ventilator Specifications
Dimensions
- Height: 53.8 In (136.7 cm)
- Width: 18.6 In (47.2 cm)
- Depth: 11.4 In (29 cm)
- Weight: 143 lbs (64.9 kg)
Bias Flow
- 0–60 liters per minute (LPM) Continuous, 15-turn control.
- Resolution: 2.5 LPM.
- Accuracy: ±10% of full scale at the following conditions: air or oxygen @ 70°F and 760 Torr.
Mean Pressure Adjust
- Approximately 3–55 cmH2O minimum range; Bias Flow dependent.
- Resolution: 0.1 cmH2O on airway pressure digital meter, 1-turn control.
- Accuracy: Non-calibrated control knob.
Mean Pressure Limit
- Automatic
Power
- At 100% power, ?P >90 cmH20 max amplitude of proximal airway pressure.
- Resolution: Graduated 10-turn locking dial, not calibrated in % power.
Frequency-Hz
- 3–15 Hz oscillator frequency.
- Resolution: 0.1 Hz on digital meter, 10-turn control.
- Accuracy: ±5% of full scale.
% Inspiratory Time
- 30–50% of oscillatory cycle.
- Resolution: +/- 1% as read on digital meter
- Accuracy: ±5% of full scale.
Set Max Paw Alarm Thumbwheel
- 0–59 cmH2O mean airway pressure.
- Resolution: 1 cmH2O
- Accuracy: Within ±2 cmH20.
Set Min Paw Alarm Thumbwheel
- 0–59 cmH2O mean airway pressure.
- Resolution: 1 cmH2O.
- Accuracy: Within ±2 cmH20.