Stryker Neptune 2 Waste Management System

24-Hour Quotes

Pricing up to 50% below the cost of new,
with service and warranty included.
Download Specifications

Stryker Neptune 2 Features

The Stryker Neptune 2 Waste Management System is a closed fluid waste and smoke evacuation system. The mobile unit collects surgical waste from operating rooms. The Neptune 2 can also capture and eliminate smoke at the same time. The waste management system help dispose of Medical OR waste while reducing exposure to the surgical fluids.

  • SealShut Technology is constantly closed during cases, manifold changes, docking and transport.
  • Systematic containment reduces fluid contact with staff and floor.
  • Permits patient-to-patient use (Manifolds are single use).
  • Clearly defines and displays low, medium and high suction and audibly alerts and visually displays when on high suction.
  • Provides two independently adjustable suction sources, one for each canister.
  • Powered IV pole for four 3000mL bags; self retracts when turned off.
  • Helps to optimize operating room turnover time,
    scheduling and volume.

Stryker Neptune 2 Specifications

Dimensions

  • Height: 16 in (40.6 cm)
  • Width: 23 in (58.4 cm)
  • Depth: 23 in (58.4 cm)
  • Weight: 95 lbs (43 kg)

Water Requirements

  • Pressure Range: 345 kPa to 827 kPa (50 to 120 psi)
  • Temperature Range: 40 °F to 110 °F (4.4 to 43.3 °C)
  • Fitting Connection: Facility source is equipped with a 3/4” Male (garden) Hose Thread (MHT) fitting and has a dedicated shutoff valve.
  • Water quality: potable tap water
  • Water Usage: Approximately 34 liters (9 gallons) per rinse cycle at default settings on standard cycle; water usage fluctuates due to selected cycle and facility flow.

Water Inlet Hose

  • Inner Diameter: 0.50 in ( 1.27 cm)
  • Length: 6ft (1.83 m)

Waste Outlet Hose

  • Inner Diameter: 1 in (25.4 cm)
  • Length: 6 ft (1.83 m)

Environmental

  • Operation Temperature Limitation: 50°F to 104°F (10°C to 40°C)
  • Operation Humidity Limitation: 30% to 75%
  • Atmospheric Pressure Limitations: 70 kPa to 106 kPa
  • Storage and Transportation Temperature: -4°F to 104°F (-20°C to 40°C)
  • Storage and Transportation Humidity: 10 to 75 %
  • Storage and Transportation Atmospheric Pressure: 50 kPa to kPa