GE Dinamap CARESCAPE V100


Download the specifications (PDF)

Vital Signs Monitor

  • Cofigurations:
    o NIBP/Rec (standard)
    o NIBP/SpO2
    o NIBP/SpO2/Temp

    • Options:
    o Nellcor SpO2
    o Massimo SpO2

    SPECIFICATIONS

    Alaris Turbo-Temp Specs
    Scale °Fahrenheit (F); °Celsius (C)
    Range
    Predictive mode Max: 41.1°C; 106.0°F Min: 35.6°C; 96.0°F
    Monitor mode Max: 41.1°C; 106.0°F Min: 26.7°C; 80.0°F
    Monitor mode accuracy ±0.1°C; ±0.2°F

    Printer Spec
    Printer type: Thermal dot array
    Resolution: 384 dots/inch horizontal
    Paper type: The paper roll used by the printer must be compatible with GE PN 770137.
    Languages printed: English, German, French, Italian, Spanish, Portuguese (Brazil and Portugal), Hungarian, Polish, Czech, Finnish, Swedish, Danish, Dutch, Norwegian, and Slovak

    Battery Specs
    Capacity: 6V; 3.3 Ahr sealed lead acid battery protected by internal auto-resetting fuse and thermal protection

    Battery Life:
    8.1 hours with a usage scenario of: NIBP determinations every 15 minutes with SpO2 and temperature active. 11.5 hours non-SpO2 versions with a usage scenario of: NIBP determinations every 15 minutes with temperature active.
    Charge time: Approximately 5 hours from full discharge when the monitor is off. Approximately 8 hours when the monitor on.

    Nellcor Oxi-Max Specifications
    Measurement Range
    SpO2 1 to 100%
    Pulse Rate 20 to 250 bpm
    Perfusion Range 0.03 to 20%
    Accuracy
    Saturation

    Adult* 70 to 100% ±2 digits
    Neonate* 70 to 100% ±3 digits
    Low Perfusion** 70 to 100% ±2 digits
    Pulse Rate
    Adult and neonate 40 to 250 bpm ±3 digits
    Low Perfusion** 40 to 250 bpm ±3 digits
    Factory Default Settings
    SpO2 HIGH Alarm Limit 100%
    SpO2 LOW Alarm Limit 90%
    Response mode 1
    (for Mode 1: Normal Response)
    SatSeconds™ 0
    Masimo SET Specifications
    Measurement Range

    SpO2 1 to 100%
    Pulse Rate 25 to 240 bpm
    Perfusion Range 0.02 to 20%
    Accuracy and Motion Tolerance
    Saturation

    Without Motion adult/pediatric* 70 to 100% ±2 digits
    Without Motion neonate* 70 to 100% ±3 digits
    With Motion adult/ped/neonate**† 70 to 100% ±3 digits
    Low Perfusion‡ 70 to 100% ±2 digits
    0 to 69% unspecified
    Pulse Rate
    Without Motion 25 to 240 bpm ±3 digits
    With Motion normal physiologic range 25 to 240 bpm ±5 digits
    Low Perfusion Performance
    0.02% Pulse amplitude Saturation (% SpO2) ± 2 digits
    % transmission >5% Pulse rate ± 3 digits
    NIBP Specifications
    Cuff Pressure Range (Normal operating range)
    0 to 290 mmHg (adult/ped)
    0 to 145 mmHg (neonate)
    Blood Pressure Accuracy (SuperSTATTM NIBP algorithm)
    Blood Pressure Accuracy (Classic and Auscultatory)
    Meets ANSI/AAMI Standard SP-10:1992 (mean error =5 mmHg, standard deviation =8 mmHg)
    Meets ANSI/AAMI Standard SP-10:2002 (mean error =5 mmHg, standard deviation =8 mmHg)
    Maximum Determination
    120 s (adult/ped)
    85 s (neonate)
    Overpressure Cutoff 300 to 330 mmHg (adult/ped)
    150 to 165 mmHg (neonate)
    Blood Pressure Range (SuperSTAT NIBP Algorithm)
    Systolic 30 to 290 mmHg (adult/ped)
    30 to 140 mmHg (neonate)
    MAP 20 to 260 mmHg (adult/ped)
    20 to 125 mmHg (neonate)
    Diastolic 10 to 220 mmHg (adult/ped)
    10 to 110 mmHg (neonate)
    Blood Pressure Range (Classic and Auscultatory)
    Systolic 30 to 245 mmHg (adult/ped)
    40 to 140 mmHg (neonate)
    MAP 15 to 215 mmHg (adult/ped)
    30 to 115 mmHg (neonate)
    Distolic 10 to 195 mmHg (adult/ped)
    20 to 100 mmHg (neonate)
    Pulse Rate Range (SuperSTAT NIBP algorithm)
    30 to 240 beats/min (adult/ped)
    30 to 240 beats/min (neonate)
    Pulse Rate Range (Classic and Auscultatory)
    30 to 200 beats/min (adult/ped)
    30 to 220 beats/min (neonate)
    Pulse Rate Accuracy ± 3.5% or 3 bpm
    GE Ohmeda® SpO2 Specifications
    Measurement Range

    SpO2 1 to 100%
    Pulse Rate 30 to 250 bpm
    Perfusion Range 0.03 to 20%
    Accuracy
    Saturation

    Adult 70 to 100% ±2 digits whichever is greater, (without motion)
    Neonate* 70 to 100% ±3 digits (without motion)
    Adult/Neonate** 70 to 100% ±3 digits (during clinical motion)
    Low Perfusion 70 to 100% ±2 digits (during clinical low perfusion)
    Pulse Rate
    Adult/Neonate 30 to 250 bpm: ± 2 digits or ± 2%, whichever is greater, (without motion)
    30 to 250 bpm: ± 5 digits (during motion)
    Low Perfusion 30 to 250 bpm ±3 digits
    Mechanical Dimensions
    Height 7.7 in (19.5 cm)
    Width 8.6 in (21.9 cm) without temperature; 10.0 in (25.4 cm) with temperature
    Depth 5.3 in (13.5 cm)
    Weight 5.4 lb (2.4 kg) including battery
    Power Requirements
    Power converter 2018859-001 universal P/N
    Protection against Class II electrical shock
    AC input Voltage 100 to 250VAC, 12VA
    DC output Voltage 12VDC at 1A
    The AC mains power adapter contains a nonresettable and nonreplaceable fuse.
    Monitor
    Protection against Internally powered or Class II when powered electrical shock from specified external power supply.
    DC input voltage 12 VDC, supplied from a source conforming to IEC 60601-1.
    Fuses: The monitor contains three fuses. The fuses are mounted within the monitor. The fuses protect the low voltage DC input, the battery, and the remote alarm output. The +5 V output on the host port connector is regulated by internal supply.
    Battery Refer to “Battery” Section

*Soma Technology, Inc. acknowledges all registered trademarks of manufactures' listed. The technical data given in this publication are for general information and is subject to change without notice.