Physio-Control Lifepak 9 Features
The Physio-Control Lifepak 9 is a defibrillator/monitor with a microprocessor. The Lifepak 9 defibration system offers a simple, user-friendly operation. The Lifepak defibrillation system features both high and low-energy selections, cardioscope, non-invasive pacing, and 3-lead and 5-lead ECG monitoring. The Physio-Control Lifepak 9 also features an adaptor that allows for hands-free defibrillation. The microprocessor determines how many joules of energy are required for defibrillating a patient. Allowing the system to readjust to low battery power or lower voltage uses, with paddles for both adult and pediatric patients.
- Microprocessor-based unit with optimal pacing
- Can operate on either AC power or internal battery power
- Built-in recorder and 3-5 lead ECG monitoring
- CRT Monitoring can display heart rate and alarms
- Front panel push buttons provide controls for most functions
Physio-Control Lifepak 9 Specifications
Dimensions
- Height: 35.2 cm (13.9 in)<//li>
- Width: 29.7 cm (11.7 in)
- Depth: 31cm (12.2 in)
- Weight: 12.7 kg ( 28 lbs)
General specifications
- Power Cord Length: 3m (10 Ft)
- Standard Paddle Electrode area: 82 square cm
- AC Input Option: 120 or 240 Vac nominal line voltage without adjustment. 50 or 60Hz, with adjustment in the SET UP menu.
- Battery type: Sealed lead-acid, 16 Vdc nominal, 3 Amp-hour
- low battery indicator: Advises operator to connect AC power.
- Battery Charge Indicator: Illuminates when the battery is charging.
- Battery Charge Time: 24 hours to full capacity. <?li>
- Service Indicator: Indicates self-diagnostic routines have detected improper operation requiring service attention.
- Power Consumption: 160 Watts maximum while monitoring with recorder ON and defibrillator charging.
ECG Monitor
- ECG Lead Selections: Std, Paddles, I, II, III
- Input: Isolated ECG via QUIK-LOOK® defibrillator paddles, FAST-PATCH® disposable defibrillation/ ECG electrodes, or 3-lead patient cable.
- Patient Cable Length: 4.0m (13 ft) total length; 3.1m cable (10 ft) with 0.9m leads (3 ft)
- Common Mode Rejection: 100dB minimum with respect to chassis ground and 65dB minimum with respect to isolated ground, when measured at 60Hz. Common mode range for patient cable input = 10 Volts peak with respect to isolated ground.
- Monitor Display Size:102mm (4 in) wide x 76mm (3 in) tall, non-fade
- Sweep Speed: 25mm/second
- Frequency Response: Non-diagnostic
- ECG Leads:1.0 to 40Hz (-3dB) <?li>
- Paddles: 2.2 to 20Hz (-3dB)
- ECG Size: Adjusts amplitude of ECG trace on monitor and strip chart recorder. <?li>
- Heart Rate Meter: 3-digit readout displays rates from 20 to 300 bpm. Heart rates outside this range do not yield valid systole tones or heart rate displays. Heart Rate Alarm: User-selectable alarm limits. Three high/low settings (150/40, 120/60, 160/90). Other limit options are available via the SET UP menu.
- 1 mV Cal: A momentary pushbutton switch on the rear panel simulates a 1 mV signal pulse to the ECG input.
- ECG Output:1 Volt/mV
Defibrillator
- Defibrillator Waveform: 5-millisecond monophasic pulse (Edmark)
- Energy Select:
- External Paddles: 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 20, 30, 50, 100, 200, 300, 360 Joules
- Internal Paddles: 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 20, 30, 50 Joules
- Charge Controls: Independent momentary push-button controls on the front panel and APEX paddle.
- Paddle Cord Length:3m (10 ft)
- Charge Indicators: Flashing lamps on the paddle and front panel pushbutton along with increasing stored energy display indicate a charge in progress. Upon full charge, the energy available is displayed on the monitor and charge completed tone sounds.
- Charge Time: Charge to 360 Joules in less than 10 seconds with a fully charged battery.
- SYNC: Synchronizes defibrillator pulse to patientgenerated QRS complex.
- SYNC Indicator: The inverted triangle marker on the displayed ECG waveform identifies the synchronizer trigger point with respect to the patient’s QRS complex.
Thermal Array Recorder
- Paper Size: 50mm wide x 30m (100 ft) <?li>
- Paper Speed: 25mm/second <?li>
- Delay: ECG prints 8 seconds after first appearing on the monitor.
- Frequency Response: Non-diagnostic. Diagnostic available via SET UP menu.
- ECG Leads (non-diagnostic): 1.0 to 40Hz (-3dB)
- ECG Leads (diagnostic): 0.05 to 100Hz (-3dB)
- Paddles: 2.2 to 20Hz (-3dB)
- Annotation: Time, date, ECG lead, ECG gain, heart rate, defibrillation parameters, test load discharges
- CODE SUMMARY Critical Event Record: Digitally stored record of ECG and device parameters
Environmental
- Atmospheric Pressure: 797mmHg to 500mmHg (-570 to 11000 ft)
- Relative Humidity: <?li>
- 0 to 95% (non-condensing) at 0 to 34°C
- 0 to 80% (non-condensing) at 35 to 45°C
- Temperature Range: 0 to 45°C operating -30 to 65°C storage