GE Marquette Mac 5000
24-Hour QuotesPricing up to 50% below the cost of new,
with service and warranty included. Download Specifications
GE Marquette Mac 5000 Features
The GE Marquette MAC 5000 is an ECG system primarily used for resting ECG analysis readings. MAC 5000
can provide advanced ECG readings for both a 12 and 15-lead analysis, interpretation, and waveforms. The EKG system comes with, a 10-inch display, wireless communications, and networking capabilities. The GE System weighs around 15 pounds making it easy to transport by itself or by a cart to the patient.
- 15 and 12 lead EKG analysis
- Offers stress testing and resting ECG
- Gender-specific criteria for identifying arrhythmia
- 10-inch color display
- Adult and pediatric interpretation
GE Marquette Mac 5000 Specifications
Dimensions (without trolley)
- Height: 3.7 in (9.4 cm) (with display closed)
- Width: 15 in (38.1 cm)
- Depth: 13.8 in (35.1 cm)
- Weight: 15 lb (6.8kg)
Operational Conditions
- Temperature: 10°C to 40°C (50°F to 104° F)
- Relative Humidity: 20% to 95% RH noncondensing
- Atmosphere Pressure: 700 to 1060 hPa
Processing
ECG Interpretation
- 12SL Adult and Pediatric ECG Analysis Program
Computerized Measurements
- 12SL™ Adult and Pediatric ECG Analysis Program
Optional
- PHI-RES & HI-RES Late Potential Analysis
Additional ECG Function
- Vectorcardiography
ECG Analysis Frequency
- 500 samples/second (sps)
ECG Storage
- 150 typical
- 200 ECGs maximum on removable media (1.44 MB, 3.5diskette)
Digital Sampling Rate
- 4000 samples / second / channel
Pre-Acquisition
- Provides 10 seconds of instantaneous ECG acquisition
AC Differential
- ±10 mV DC offset: ±320 mV
Resolution
- 4.88 µV/LSB @ 500 sps
Frequency Response
- –3 dB @ 0.01 to 150 Hz
Common Mode Rejection
- >140 dB (123 dB with AC filter disabled)
Input Impedance
- >10M@ 10 Hz, defibrillator protected
Patient Leakage
- <10 µA
Pace Detect
- Orthogonal LA, LL and V6; 750 µV @ 50 µs
Special Acquisition Functions
- Disconnected lead detection
- Electrode impedance
- Excessive AC noise
- Baseline wander and muscle tremor messages
Heart Rate Meter
- 30 to 300 BPM ± 10% or 5 BPM whichever is greater. Heart rates outside this range will not be displayed
Display
Display Type
- 264 mm (10.4 in) diagonal graphics backlit AM LCD
Display Resolution
- 640 x 480 pixels with waveform enhancement
Display Data
- Heart rate
- Patient name
- ID
- Clock
- Waveforms
- Lead labels
- Speed
- Gain and filter settings
- Warning messages
- Prompts and help messages
Writer
Writer Technology
- Thermal dot array
Writer Speeds
- 5 mm/s
- 12.5 mm/s
- 25 mm/s
- 50 mm/s (same as display)
Number of Traces
- 3
- 6
- 12
- 15
- user selectable (same as display)
Writer Sensitivity / Gain
- 2.5, 5, 10, 20, 10/5 (split calibration) mm/mV (same as displayed)
Writer Speed Accuracy
- ±2%
Writer Amplitude Accuracy
- ±5%
Writer Resolution
- Horizontal 1000 dpi @ 25 mm/s
- 200 dpi vertical
Paper Type
- Thermal, Z-fold, perforated, fan fold, 300 sheets/pack
Paper Size
- A size: 214.63 mm x 280 mm (8.45 in x 11 in)