CSZ Hemotherm 400CE Sub Zero Cooler
24-Hour QuotesPricing up to 50% below the cost of new,
with service and warranty included. Download Specifications
CSZ Hemotherm 400CE Features
The CSZ Hemotherm 400CE dual reservoir heater/cooler is used to lower, maintain, or raise the temperature of the water flowing through a blood oxygenator/heat exchanger that is used to cool or warm blood during cardiopulmonary bypass procedures lasting 6 hours or less. This Sub Zero cooler may also be used with a hyper/hypothermia blanket under the patient to provide warming through conductive
heat transfer.
- Temperature-controlled water in a full range of controllability between 37.4°F and 107.6°F.
- High water flow rates at low pressures.
- Able to maintain normal body temperature, particularly during lengthy surgical procedures with a hyper-hypothermia blanket.
- Optional remote control model 414CE.
CSZ Hemotherm 400CE Specifications
Dimensions
- Height: 32” (81 cm)
- Width: 22” (56 cm)
- Depth: 22” (56 cm)
- Weight: 198 lbs ( 89.8 kg)
Reservoir Capacity
- Cool: 8 qts (7.6 liters)
- Heat: 6 qts (5.7 liters)
- Reservoir Fluid: Distilled Water Only
- Reservoir Opening: Easy-to-fill
- Reservoir Construction: Plastic
Flow Rate
- 13 Liters/Minute (3.5 GPM) through self-sealing quick-connect fittings
- 15 Liters/Minute (4 GPM) with flow through quick connect fittings
Maximum Pressure
- 12.5 PSI – Heat Exchanger and Blanket
Connections
- 1 Set 1/2” quick connect fittings for Heat Exchanger
- 1 Set 1/8” quick connect fittings for Blanket
- 1 Set 1/4” quick connect fittings for draining (on unit)
- 1 Additional set of 1/2” quick connect fittings included for tubing
Safety
- Secondary Hi-Limit Thermostat: 114.8°F ± 32.9°F (46°C ± 0.5°C)
- Primary Hi-Limit Thermostat: 111.2°F ± 32.9°F (44°C ± 0.5°C)
- Low Limit Thermostats: 35.6°F ± 32.9°F (2°C ± 0.5°C)
Electrical Characteristics
- 115VAC, 50/60HZ, 16 Amps
- 230VAC, 50/60HZ, 9 Amps
Outlets Required
- 115VAC – 20 Amps, AC grounded hospital grade plug
- 230VAC – 15 Amps, AC grounded with a European plug
Circuit Breaker Rating
- In power switch
- 115VAC, 20 Amps
- 230VAC, 15 Amps
- Mains Supply Isolation
- Two-Pole Mains Power Switch