Stryker Altrix Temperature Management System
24-Hour QuotesPricing up to 50% below the cost of new,
with service and warranty included. Download Specifications
Stryker Altrix Features
The Stryker Altrix is a temperature management system that is used to regulate patient temperatures to maintain patient control within ±0.1°C. The Altrix utilizes 3 water circuits and maximizes performance to achieve the target temperature quickly and safely. The target temperature is achieved with non-invasive and non-adhesive body wraps redesigned to increase patient surface contact and air gaps. This Stryker temperature management system comes with pre-configured settings for a wide variety of hypothermia, hyperthermia, or other temperature management applications.
- Three water circuits with self-sealing ports to reduce water spillage. Each circuit can be individually monitored.
- User-friendly graphical user interface.
- Easy maneuverability with an ergonomic design.
- Removable translucent water reservoir that allows uninterrupted performance while refilling.
Stryker Altrix Temperature Management System Specifications
Dimensions
- Height: 42.5” (107.9 cm)
- Width: 15” (38.1 cm)
- Depth: 23” (55.4 cm)
- Weight (empty): 150 lbs (68 kg)
- Reservoir Capacity: 1.3 gal (5 L)
Water Temperature
- Control Setting Range: 39.2°-104.0° F (4.0°-40.0° C)
- Control Accuracy: ±0.3° C (4.0°-40.0° C)
- Display Measurement Accuracy: ±0.2° C (4.0°-40.0° C)
- Display/Resolution: 0.1° C
- Default: 104.0° F (40.0° C)
Water Temperature
- Control Setting Range: 39.2°-104.0° F (4.0°-40.0° C)
- Control Accuracy: ±0.3° C (4.0°-40.0° C)
- Display Measurement Accuracy: ±0.2° C (4.0°-40.0° C)
- Display/Resolution: 0.1° C
- Default: 104.0° F (40.0° C)
Patient Temperature
- Control Setting Range: 89.6°-100.4° F (32.0°-38.0° C)
- Control Accuracy: ±0.1° C (32°-38°C)
- Measurement Accuracy: ±0.3° C (25.0°-45.0° C) ±0.4° C (0.0° C-24.9° C, 45.1° C-50.0° C)
- Display/resolution: 0.1° C
- Display Range: 32.0°-122.0° F (0.0°-50.0° C)
- Default Setting: 98.6° F (37.0° C)
Controller
- Heater Capacity Max: 500 watts
- Circulating Fluid: Sterile distilled water
- Battery: 9V lithium
- Alarm Tone Range: 75-85 dBA per standard; IEC 60601-1-8
- Water flow rate in each hose port typical: 1.2 lpm
- Refrigerant Type: R134a
- Power Cord Length: 14 to 15 feet (4.2-4.5 meters)
- Clinical Thermometer: Direct mode