B. Braun Vista Basic Infusion Pump
24-Hour QuotesPricing up to 50% below the cost of new,
with service and warranty included. Download Specifications
B. Braun Vista Basic Infusion Pump Features
The B.Braun Vista Basic is a volumetric infusion pump intended to provide an accurate and continuous flow of parenteral and enteral fluids. The standard delivery rate ranges from 0.1 to 999.9 ml/h with intrafix and 0.1 to 800 ml/h with Vista Pump Set. This unit is operated via a membrane keypad and equipped with a liquid crystal display to show the delivery rate and operating support to the user. As a workhorse in the outpatient market, it offers a rugged and reliable design, all relevant therapies such as continuous, piggyback, and program mode, and a selection of cost-effective, straight-line gravity sets.
- The robust design of mechanical components ensures maximum durability.
- Linear peristaltic pump.
- Can be operated as a stand-alone unit or integrated into an intensive care unit.
- The battery provides at least 3 hours of operation at the highest delivery rate.
- Easy to use, teach, and learn.
- Using a drop sensor eliminates the need for pre-setting a VTBD in continuous mode and senses upstream occlusions and empty solution containers.
B. Braun Vista Basic Specifications
Dimensions
- Height: 9.4 in (24 cm)
- Width: 5.5 in (14 cm)
- Depth: 7.8 in (20 cm)
- Weight: 6.8 lbs (3.1 kg)
Mechanism
- Linear peristaltic volumetric infusion pump
Power
- VAC: 100-120 VAC
- Hz: 50-60 Hz
- Watts: 13 Watts
- Leakage Current: Max 30µA
Battery
- Type: Nickel-cadmium (rechargeable)
- Amp Hours: 1.2Ah
- Volts: 7.2V
- Operating Time: ˜3.5 hours at 125 mL/hr
- Charge Time: Approx. 16 hours
Infusion Rates
- Intrafix Tubing: 0.1 to 999.9 mL/h
- Vista Tubing: 0.1 to 800.0 mL/h
- KVO Rate: 1 mL/h or selected rate, whichever is lower
- Flow Rate Accuracy: ±5% measurement time 8h5, temp. 71.6°F
Alarms
- Air-in-Line, Low Battery, Plug In Now, Bag Empty, Downstream Occlusion, Door Open, Pump Stopped, KVO, Unit Malfunction
Air-in-Line Alarm
- Air bubble size 0.3mL (Can be set from 0.01 to 0.3 mL in service) or 1.5mL/hr (Can be set from 0.5 to 3.5 mL/hr in service program)
Modes/Features
- Continuous and Piggyback Modes, Program Mode for variable infusions, Ramp/Taper Mode for TPN infusions and Delay A Mode for postponing infusion start and finish
Memory
- User option to retain last Continuous and Piggyback Mode data upon power up. Permanent data retention in Program and Ramp/Taper Modes after the pump is turned off
Operating Temperatures
- Operating: 64.4° F to 95° F, 10.15 PSI to 15.37PSI
- Non-Operating: -13° F to 131° F, 10.15 PSI to
- 15.37 PSI