GE OEC 6800 MiniView

Quotes in less than 24 hours Download Specifications

OEC Mini C-arm Features

The GE OEC 6800 MiniView is a perfect fit for sports medicine, orthopedics, and veterinary medicine. The OEC 6800 is the first mini c-arm to use 1k2 imaging and a touchscreen interface. This OEC c-arm provides many of the components of a full-sized system at a fraction of the cost.

Movements

Together the articulating arm, rotating arm, and the OEC C-arm provide maximum flexibility in positioning the system for each procedure. It makes elevation and horizontal positioning easy. The C-arm can be elevated so that the isocenter is in the range of 23-52″ (58-132 cm) with the image intensifier down. In addition to the elevation, the articulating arm pivots horizontally at both ends. OEC’s MiniView can be angled by rotating the C-arm on the rotation arm. The range of movement is over 200 degrees. The C-arm can also orbit in its cradle through an arc of 115 degrees.

Touchscreen Controls

The right monitor provides a touchscreen interface to interact with the MiniView system. Using the touchscreen on the equipment you can; enter and view patient information, view and process images, configure the X-ray switches, set imaging parameters, and annotate images.

Workstation Keyboard

The OEC 6800 MiniView system keyboard controls the navigation, text entry, image processing, and other function keys. There are two versions of the MiniView system keyboard that are available which include text and icons only. The layout and functionality of each keyboard is the same, but the labeling is different.

Laser Aimer

The GE OEC 6800 MiniView is equipped with a laser aimer to help position the anatomy in the center of the intensifier. To activate the laser aimer by pressing the button on the front surface of the C-arm. The Laster light is a cross that is projected toward the image intensifier.

Imaging

At the system startup, the MiniView 6800’s imaging capabilities are immediately available from the footswitch or hand switch. You can complete an exam by turning on the power, entering patient information, and pressing the X-ray switch. The system automatically selected the best combination of X-ray imaging parameters to produce an optimal image. For each new exam, common default settings are automatically applied to the Workstation. You can override the default settings to manually set the edge enhancement, noise filtering, contrast, and brightness values. These settings will automatically be reset for each new exam or scheduled exam. How many images can be stored on the internal system? 18 images can be stored on the CP (Clinical platform) system, and 400 images can be stored on the HP (hospital platform) system. The GE 6800 MiniView system supports a variety of hard copy and archive devices. Depending on the device options installed on your system you can send images, patient information, and the dose summary to storage and hard-copy devices.

Hardcopy Devices
  • Instant film/paper printer
  • DICOM printer
Storage Devices

Available storage options include:

  • Jaz Disk – removable high capacity disk
  • Floppy disk
  • DICOM storage

GE 6800 C-arm Specifications

Base Dimensions

  • Height: 65” (168.1 cm)
  • Width: 27.75” (70.7 cm)
  • Depth: 27.5” (70.7 cm)

C-Arm Extension

  • Maximum Front: 54.5” (138.4 cm)
  • Max. Height I.I. Up: 71.5” (181.6 cm)
  • Min. Height I.I. Down: 5.5” (14 cm)

Transport Position

  • Height: 65” (165.1 cm)
  • Min. Width: 33.5” (85.1 cm)
  • Min. Depth: 32.5” (82.6 cm)

Stability

  • Positive stability on grade up to 5 degrees in any position.
  • Positive stability on grades up to 10 degrees in the transport position.

Imaging Area

  • Depth: 17.7” (44.95 cm)
  • SID: 17.7” (44.95 cm)
  • Free Space: 13.5” (34.27 cm)
  • Skin Spacer: 3.9” (10 cm)

Arm Movements for Medical Procedures

  • Orbital Travel: 118 degrees
  • Wig-Wag: 135 degrees
  • Total Rotation ±200 degrees

Laser Aimer

  • Dual Fan Beam Laser Diodes
  • Laser Class IIIA
  • Power Output: 5mW max.
  • Wavelength: 650 nM
  • Fan Beam Coverage: 7.87” (20 cm)

Image Intensifier

  • Normal Nominal Entrance Field Size: 5.9” (15 cm)
  • Mag Nominal Entrance Field Size: 3.94” (10 cm)
  • Supply Input Voltage: 24 ± 1 VDC
  • Input X-ray Exposure: 1 R/minute max.
  • Central Resolution: 66 Lp/cm (typical)
  • Contrast Ratio: 30:1 (typical)
  • Conversion Factor: 180 cd/m2 (typical) mR/S
  • DQE: 42%

Fluoroscopic Accuracy

  • µ?: ±5%
  • kV: ±5%

X-Ray Source Assembly

  • Type: Fixed anode, tungsten filament
  • Target Angle: 14 degrees
  • Focal Spot: 0.05 MM
  • Inherent Filtration: 1.0 mm Al Equivalent
  • Total Filtration: 2.5 mm Al Equivalent
  • Beam Quality: Minimum HVI, 2.8 mm Al @ 80 kV
  • Leakage Technique Factors: 80 kV and 160 µA

Power

  • Line Frequency: 60/50 Hz
  • 100/120 VAC (±10%): 6 Amps
  • 200/220/230/240 VAC (±10%): 4 Amps
  • Max. Non-Destructive voltages that can be connected to signal I/O ports: 7 Volts DC
  • Room I/O Port: +24V relay closures, +5TTL logic levels

Video Signal

  • Composite video, EIA RS170A 60 Hz, 525 lines (International CCIR 50 Hz, 625 lines)

Environmental Requirements

  • Operating Temp.: 50 to 95°F (10 to35°C)
  • Storage Temp.: +14 to 131°F (-10 to+55°C)
  • Operating Altitude: 8,202 ft. (2.5 km)
  • Storage Altitude: 15,000 ft. (4.6 km)
  • Operating Humidity: 20 to 80% non-condensing
  • Storage and Transport Humidity: 20 to 80% on-condensing.
  • Shock and Vibration: 1G at 200 Hz for 2 hours.)

Image Processing and Storage

  • CP Image Storage: 18
  • HP Image Storage: 400
  • Frame Integration: Yes
  • Image Matrix Size: 1000 x 1000 x 16
  • Last Image Hold: Yes
  • Video storage digital memory PCMCIA drive, hard disk, floppy disk, SVHS VCR(optional)

Monitors

  • Dual Monitors
  • Monitor Size: 16” (40.64 cm)

Banner

Questions about our Mini C-Arms?

Soma Tech Intl offers a wide range of mini C-arms. If you have any questions about our mini C-arms or need a quick quote, call us and one of our knowledgeable sales representatives will help you.

Looking to sell your mini C-arms? Soma ensures your facility gets top dollar for your equipment and also gives the option for trade-in credit.

1-800-GET-SOMA
Contact Us