Philips Tempus Pro Features
The Philips Tempus Pro is an advanced vital signs patient monitor with multiple capabilities. The Tempus Pro provides a full range of vital signs monitoring as well as ultrasound and video laryngoscope capabilities. It has a large color display with multiple waveform configurations. The Tempus Pro system is lightweight (6.2 lbs) and highly durable with an IP66 rating.
- 10-hour internal battery life.
- Fully integrated communications capability enables the transmission of all medical and vital signs data, voice, and video transmission in real-time using military radios.
- Easy-to-use touchscreen display that can be used with a gloved hand.
- Displays ultrasound and video laryngoscope images on the color display.
- Wi-Fi, Bluetooth, GSM, and GPS.
Philips Tempus Pro Specifications
Dimensions
- Height: 8.4” (21.4 cm)
- Width: 10.1” (25.9 cm)
- Depth: 3.9” (10 cm)
- Weight: 6.2 lbs (2.8 kg)
Display
- Color 6.5 ” VGA screen
- 130 klux daylight readable display
- Multiple user-selectable display formats
- High-contrast mode
- NVG friendly
- The device can be set to an appropriate viewing angle with the integral foot
- The rear of the Tempus Pro houses the RapidPak clip which enables easy deployment and storage of sensors and cables
ECG Monitor
- 3/5 Lead monitoring via standard snap-on electrodes
- Heart rate range: 30-300 BPM
- 12 Lead acquisition / Interpretation
- Input impedance: >100 MO
- Dynamic range: ±5 mV ac
- Accuracy: ±3%
- DC offset: ±300 mV dc
- Frequency Response: 0.05 Hz to 175 Hz ±3dB
- Sample rate: 500 Hz
- Common mode rejection: 95 dB minimum, additional filters include mains, muscle, and low and high pass
- Arrhythmia monitoring
Impedance Respiration
- Range: 3 – 150 RPM
- Accuracy: ±2 RPM or ±2% whichever is greater
Non-Invasive Blood Pressure
- Accuracy: ±3 mmHg or ±2% (whichever is greater)
- Adult range: 20 – 260 mmHg
- Child range: 20 – 250 mmHg
- Neonate range: 20-140 mmHg
- Available Cuffs: Neonate disposable 1-5, Infant, Child, Child long, Adult, Adult Long, Large Adult, Large Adult Long, Thigh
Pulse Oximetry
- Range: 1 – 100%
- Accuracy (adults/child): no motion or low perfusion ±2 digits 70-100%, motion ±3 digits 70-100%
- Accuracy (neonate): motion, no motion, and low perfusion ±3 digits 70-100%
- Perfusion index: 0.02-20%
- Response: <1 second delay
- Masimo SET Rainbow technology
- Pleth Variability Index (PVI) v
- Pulse Rate Range: 25 – 239 BPM
- Pulse Rate Accuracy (all ages): no motion =3 digits, motion =5 digits
- Haemoglobin
- Range: 0 – 25 g/dl
- Accuracy (Adults/Infants/Paediatrics): 8 – 17 g/dL ± 1 g/dl
- Methaemoglobin
- Range: 0 – 99.9%
- Accuracy (Adults/Infants/Paediatrics Neonates): 1 – 15% ± 1%
- Carboxyhaemoglobin
- Range: 0 – 99.9%
- Accuracy (Adults/Infants/Paediatrics): 1 – 40% ± 3%
- Total oxygen content: Range 0 – 35ml of O2/dl of blood
Capnometry
- Respiration Rate
- Range: 1 – 149 BPM
- Accuracy: 0-70 BPM ±1 BPM, 71-121 BPM ±2 BPM, 122-149 BPM ±3 BPM
- Microstream ETCO2
- Range: 0 – 150 mmHg
- Flow rate: 50 ml/min -7.5 + 15 ml/min
- Uses Oridion Microstream technology
- Accuracy: 0-38 mmHg ±2 mmHg, 39-150 mmHg ±5% of reading +0.08% per 1 mmHg over 38 mmHg
Contact Temperature
- 2-channel YSI 400 series compatible
- Measurement range: 20-45 °C
- Resolution: 0.1 ºC (0.2 °F)
- Accuracy: ±0.1 ºC
Integral Digital Camera
- Color 3.2M pixel camera
- Takes still pictures or video using the H264 algorithm (bandwidth dependent)
- Images are included in the patient record
Invasive Pressure
- 2 channels, 5 µV/V/mmHg
- Response: 0-20 Hz (-3 dB)
- Filters: 50/100/150 Hz & 50-60 Hz notch
- Range: -99 – 310 mmHg
- Accuracy: ±2% or ±2 mmHg
- Bridge: 180 O minimum
- Expandable up to 4 channels via USB module
Military Trauma Record
- Electronic trauma record (TCCC, AF 3899Lix, etc.)
- User-friendly interface
- Semi-automatic patient record completion
- Operable with a gloved hand
- Data output is compatible with the military AHLTA-T records system
- Record can be passed from device to device to accompany the patient through the echelons of care
- Data can be output as. PDF for attachment into electronic patient record systems
- The record can be transmitted over military radios for real-time decision support
Ultrasound
- Optional Interson Ultrasound Probes General Purpose 3.5 MHz and Line
Placement 7.5 MHz
Video Laryngoscopy
- Optional Karl Storz C-MAC Video Laryngoscope Imager and Single-Use Blades