Bear Cub 750 PSV
24-Hour QuotesPricing up to 50% below the cost of new,
with service and warranty included. Download Specifications
Bear Cub 750 PSV Features
The Bear Cub 750 PSV is an infant ventilator that can accommodate patients that weigh from 1 pound (500 grams) to 66 pounds (30 kg). The Bear Cub was designed to support critical care of neonatal and pediatric patients with compromised lung function. The System offers integrated synchronized ventilation along with pressure, flow, and volume monitoring. The Bear Cub offers three different ventilation modes from assisted control, SIMV/IMV, and CPAP.
- Pneumatic Ventilator
- Gas Inlet Conditioning
- Accurate blending of gases in the O2 blender
- Flow Control Valve
- Auxiliary Gas Outlet
- Pressure Control System
- Flow Sensor
Bear Cub 750 PSV Specifications
Ventilator Dimensions
- Height: 11 in
- Width: 13.5 in
- Depth: 10 in
Graphic Display Dimensions
- Height: 9.5 in
- Width: 13 in
- Depth: 2.5 in
- Weight: 5.5 lbs
Pedestal Stand Dimensions
- Height: 40.5 in
- Width: 24 in
- Weight: 9 lbs.
Compressor Dimensions
- Height: 36 in
- Width: 22 in
- Depth: 21.5 in
- Weight: 11o lbs
Environmental
- Temperature
- Storage and Shipping -40 to 158 ºF (-40 to 70 ºC)
- Checkout and Operating 50 to 104 ºF (10 to 40 ºC)
- Altitude
- Checkout, Operating, Transport and Storage 0 to 9,000 feet (14.7 to 10.5 PSIA/760 to 543 mm Hg)
- Humidity
- Storage and Shipping 0 to 99% Relative Humidity — Non-condensing
- Checkout and Operating 0 to 95% Relative Humidity — Non-condensing
Controls
- Mode: Flow Cycle AC, AC, SIMV/IMV, SIMV/PSV, Flow Cycled SIMV CPAP/PSV
- Rate: 1 to 150 bpm
- Inspiratory Time: 0.1 to 3.0 sec
- PEEP: 0 to 30 cmH2O
- Inspiratory Pressure: 0 to 72 cmH2O
- Manual Breath: x1
- Assist Sensitivity: 0.02 to 5.0 L/min
- Over Pressure Relief: 15 to 75 cmH2O
- % O2: 21 to 100%
- Apnea Alarm (selectable): 5, 10, 20 or 30 sec
Monitors
- Breath Rate: 0 to 255 bpm
- Patient Initiated Indicator: LED
- Minute Volume: 0 to 30.0 L/min
- Tidal Volume: 0 to 500 ml (Inspired and Expired)
- %Tube Lea:k 0 to 100%
- Inspiratory Time: 0.1 to 3.10 sec
- Expiratory Time: 0 to 99.9 sec
- I:E Ratio: 9.1 to 1:9.9
- Peak Inspiratory Pressure: 0 to 99 cmH2O
- Mean Airway Pressure: 0 to 75.0 cmH2O
- PEEP: 0 to 30.0 cmH2O
- Air Pressure: 0 to 100 psig
- O2 Pressure: 0 to 100 psig
- Proximal Airway Pressure: 10 to 100 cmH2O (Gauge)
- Hourmeter: 0 to 99,999 hours
- Test: Push button
Alarms / Limits
- High Breath Rate: 3 to 255 bpm
- Low PEEP/CPAP: -5 to 30 cmH2O
- Low Inspiratory Pressure: 0.25 (High Pressure Limit – Low PEEP/CPAP) + Low PEEP/CPAP
- Patient Circuit: LED On/Off
- Failed to Cycle: On/Off
- Low Gas Supply: LED On/Off
- Apnea: LED On/Off
- Pressure Settings Incompatible: LED On/Off
- Prolonged Inspiratory Pressure: LED On/Off
- Flow Sensor: LED On/Off
- Low Battery: LED On/Off
- Alarm Silence: 60 sec
- Visual Reset: Push Button
- High Pressure Limit: 10 to 75 cmH2O
- Line Power: Green/Red LED
Electrical
- Voltage
- 100 V – 80 to 110 VAC
- 120V – 96 to 132 VAC
- 220V – 176 to 242 VAC
- 240V – 192 to 264 VAC
- Current
- 100V – 1.0 A maximum
- 120V – 1.0 A maximum
- 220V – 0.5 A maximum
- 240V – 0.5 A maximum
- Frequency
- 100V – 50.60 Hz
- 120V – 50/60 Hz
- 220V – 50/60 Hz
- 240V – 50/60 Hz
- Fuses
- 100/120V – T o.5 A, 5 x 20 mm
- 230/240V – T 0.5 A, 5 x 20 mm
- (For 220V operation, set the power entry module seclector switch to 230V setting)