GE Giraffe Stand-alone Infant Resuscitation System
24-Hour QuotesPricing up to 50% below the cost of new,
with service and warranty included. Download Specifications
Giraffe Stand-alone Infant Resuscitation System Features
The GE Giraffe Stand-alone Infant Resuscitation System can bring life-saving equipment that can revive an infant. The potable GE Giraffe Resuscitation System can be brought to the newborn in need, from labor and delivery, NICU, ICU, and the Nursery. The system offers full resuscitation capabilities along with SpO2 and oxygen level Monitoring at the same time. The System can support both t-pieces and bag-and-mask configurations for resuscitation. This item can easily be mounted onto the GE Giraffe, GE Giraffe OmniBed, and GE Panda for convenience.
- Both T-piece and Bag-and-Mask resuscitation configurations
- Effective airway management, with fully integrated suction
- Monitored delivery of positive pressure during ventilation
- Blended Air/O2 gas delivery to meet the clinical needs throughout the stabilization period
- Preparedness/Convenience
- Accepts both wall and tank gas inputs
- Priority valve conserves tanks so they are available for transport
Giraffe Stand-alone Infant Resuscitation System Specifications
Dimensions
- Depth (knobs to fittings): <8.7 in (22 cm)
- Height <14.6 in (37 cm)
- Width <11.8 in (30 cm)
- Weight Approx. 15 lbs (7 Kg)
Environmental Specifications
- Operating temperature: 18 to 41°C
- Operating humidity: 0 to 90%
- Storage temperature: –25 to 60°C
- Storage humidity: 0 to 95%
- Storage pressure: 50 to 106 kPa
System Characteristic
- Input source: Hospital pipeline or cylinder with regulator
- Input pressure: 40-75 psi (275-600 kPa) (Air/O2 fittings)
- Minimum input: 70 pm
- Vacuum range: 0-150 mm Hg
- Vacuum gage accuracy: ±5% of full scale
- Flow capacity: 15 pm
- Flow range: 0-15 lpm
- Flow accuracy: –5 lpm ±½ lpm
- 5-10 lpm ±1.5 lpm
- 10-15 lpm ±2 lpm
Airway Pressure Manometer
- Range: –10 to 80 cmH2O
- Accuracy: ±5% of full scale
Air/O2 Blender
- Range: 21-100% O2
- Accuracy: ±5% of full scale
User Control Settings
- Air/O2 flowmeter set range: 0-15 lpm
- Suction regulator set range: 0–150 mmHg
Recommended Body Weight Range
- Up to 22 lbs (10kg)
Approximate Operating Time
- @ 5 L/min: 121 min
- @ 10 L/min: 74 min