Newport E100i
24-Hour QuotesPricing up to 50% below the cost of new,
with service and warranty included. Download Specifications
Newport E100i Features
- Transportable Ventilator
- A universal ventilator that is pneumatically powered and electronically controlled.
- Precise calibrated controls for respiratory rate, flow, inspiratory time, PEEP/CPAP and oxygen concentration.
- Adjustable pressure relief valve to prevent excessive high pressure or to implement pressure limited ventilation.
Newport E100i Specifications
General
- FIO2 0.21 to 1.0 ± 3%
- Ventilator Modes A/C and SIMV (volume or pressure limited) spontaneous.
- Inspiratory Pressure
- Range 0 to 100 cm/H2O
- Trigger Level -10 to +25 cm/H2O
- Inspiratory Time 0.1 to 3.0 seconds
- Respiratory Rate 1 to 120 breaths per minute
- Tidal Volume 10 cc to 2500 cc
- Mechanical Flow 0 to 100 L/minute (0 to 1.6 L/second)
- Continuous Flow 0 to 55 L/min. On/Off switch and flowmeter
- PEEP/CPA 0 to +25 cm/H2O
- Nebulizer On/Off switch
- Manual Breath Independent from ventilator function
- Battery (optional) Gel cell. Up to four hours use. Four hours recharge. Recharges fully after 18 hours.
- Safety Pressure Relief 0 to 100 cm/H2O
- Dimensions Height: 10.5”. Width: 9.5”. Depth: 6.5”. Weight: 13 lbs.
- Power Requirements 100 to 120 VAC 50/60 Hz, 200 to 240 VAC 50/60 Hz. Fuse protected.
- Hospital grade plug.
- Gas Supply Oxygen 35 to 70 psi/50 psi normal. Air 35 to 70 psi/50 psi normal
- Internal Compressible
- Volume 0.20 liters
Alarms
- Visual/Audible High pressure, low pressure, spontaneous/apnea, 15 to 30 seconds delay. Battery – 20 minutes remaining.
- Visual Only Inspiratory time to long
- Audible Only air/oxygen mixer supply source failure. Electrical power failure
- Alarm Silence 55 seconds