Welch Allyn CP 150 Features
The Welch Allyn CP 150 is an EKG system that can also measure spirometry. The CP 150 comes with a 7-inch color touchscreen display with a keyboard. This Welch Allyn system can easily provide full-page ECG readings with the touch of a button. With options for 3, 6, or 12-lead EKG options. The Resting electrocardiograph system can store up to 100 readings on its internal memory.
- 3, 6, and 12-lead EKG readings
- 7-inch color touchscreen display
- Lightweight battery opened system
- Can store up to 100 readings
- Offers spirometry measurements
- USB and ethernet connectivity to manage patient data
Welch Allyn CP 150 Specifications
Dimensions
- Length: 15 in (38 cm)
- Height: 14.1 in (35.8 cm)
- Width: 5.4 in (13.6 cm)
- Weight: 11.7 lb (5.3 kg) (with the battery)
Display
- Type: TFT, 18 cm (7 in) color touchscreen
- Resolution: WVGA, 800 x 480
Printing
- Thermal Paper: Z-fold 21 cm (8.25 in) x 28 cm (11 in) x 200 sheets
- Thermal Paper (internal): Computer-controlled dot array, 8 dots/mm
- Thermal chart paper speed: 10, 25, or 50 mm
- GainSettings:
- Auto ECGs: 2.5, 5, 10 or 20 mm/mV, AUTO
- Rhythm ECGs: 10, 25 or 50 mm/mV
Lead configurations
- Standard, Cabera
interpretation algorithm
- MEANS (Modular ECG Analysis System)
Environmental Operating Conditions
- Temperature 10° – 40° C (50° – 104° F)
- Relative humidity 15 – 95% noncondensing (30 – 70% for printing)
- Atmospheric air pressure limits 700 – 1060 hPa (525 – 795 mmHg)
Environmental Storage Conditions
- Temperature -20° – 50° C (-4° – 122° F)
- Relative humidity 15 – 95% non-condensing
- Atmospheric air-pressure limits 700 – 1060 hPa
Other Specifications
- Input impedance >2.5 MO
- Patient leakage <10 µA (normal condition), <50 µA (single fault condition) <?li>
- Dynamic range AC differential +/-5 mV, DC offset +/-300 mV
- Resolution 2.5 µV
- A/D conversion 16 bits
- Heart rate ranges from 30 to 250 bpm
- ECG storage (in test directory) At least 100 ECG tests
- Frequency range from 0.3 to 150 Hz
- Digital sampling rate 8000 samples/second/channel for data acquisition 500/1000 samples/second/channel for adult/pediatric ECG data analysis and storage
- Pacemaker detection ANSI/AAMI EC11
- Lead detection Lead off and noise
- Power requirement Universal AC power supply ~110–240 V, ~50/60 Hz, 1.5 A maximum
- AC fuses Time-lag type, 2.0-amp 250-V rating, Little fuse 0218002P or equivalent