Puritan Bennett Achieva PSO2 Features
The Puritan Bennett Achieva PSO2 ventilator provides continuous respiratory support for pediatric to adult patients with respiratory insufficiencies or failures, in a home, institution, or portable settings. The Puritan Bennett Achieva ventilators have a compact design, are lightweight, and use portable oxygen and power sources, which make these ventilators highly portable.
- Wide range of delivery volumes, inspiratory times, and breath rates.
- Magnetic latched door panel to prevent tampering and accidental resetting.
- Support for pediatric and adult patients.
- Use in home, institutional, and portable settings.
Puritan Bennett Achieva PSO2 Specifications
Dimensions
- Height: 10.75” (27.3 cm)
- Width: 13.3” (33.8 cm)
- Depth: 15.6” (39.6 cm)
- Weight: ~32 lb (~14.5 kg)
Type of Ventilator
- Type: Positive Pressure Volume Ventilator
- Motor: 3-phase brushless motor
- Pump: Piston, 50ml to 2200ml tidal volume capacity
- Protection Against Electrical Shock: Class 1
Power
- Power Line: 100 to 240 VAC, 50 to 60 Hz, 2 Amps
- External DC Power
- 11.6 to 32.0 VDC (24 VDC optimal)
- Operating Time: At least 19 hours under normal load, and 5 hours and 30 minutes under a heavy load 24 VDC or 12 VDC, 32 Ah (Amp-hour)
- Internal Battery
- 24 VDC (nominal)
- Operating Time: At least 4 hours under normal load, and 1 hour under heavy load
- Gel cell, sealed Lead Acid
- Lithium Battery
Settings
- Volume (Vt): For assist/control 50ml to 2200ml in 10 ml steps. for SIMV 50ml to 1750ml in 10 ml steps accurate to ± 10 ml for 50-100 ml and ±10% (max 75 ml) for 100-2200 ml of the setting to a maximum delivered volume error of 75 ml.
- Inspiratory Time (Ti): 0.2 to 5.0 seconds in increments of 0.1 seconds. Accurate to ±10%
- Sensitivity (Sens.) Flow: 3 to 25 LPM in 1 LPM increments. Accurate to ±2.0 LPM. Off or 1 to 15 hPa in 1 hPa increments. Accurate to ±2.5 hPa.
- Breath Rate (f): 1 BPM to 80 BPM in steps of 1 BPM. Accurate to ±10% of 1 BPM whichever is greater.
- Pressure (P): 0 to 50 hPa in 1 hPa increments. Accurate to ±2.5 hPa of the setting once the pressure reaches the setting.
- PEEP: 0 and 3 to 20 hPa in 1 hPa in 1 hPa increments. Accurate to ±2.5 hPa
- Low Pressure: 1 to 59 hPa in increments of 1 hPa. Activates within ±2.5 hPa
- High Pressure: 2 to 80 hPa in increments of 1 hPA. Activates within ±2.5 hPa
- Flow Acceleration: OFF or ON (Inspiratory Flow = 180 lpm
- Expiratory Trigger: 15% to 55% in increments of 10%. Accurate to ±15% at 15%, to ±5% at 25%-55%
- O2 Level (O2%): 21% to 100% for tidal volumes greater than or equal to 100 ml, 21% to 70% for tidal volumes less than 100 ml in 2% increments.