Drager Narkomed GS
24-Hour QuotesPricing up to 50% below the cost of new,
with service and warranty included.
Drager Narkomed GS Features
The Drager Narkomed GS anesthesia machine is a dependable solution for surgical centers and hospitals, offering essential features to support safe and effective anesthesia delivery. With an intuitive interface, built-in safety mechanisms, and a compact design, this system provides anesthesiologists with precise control over ventilation and gas delivery. Designed for durability and ease of use, the Narkomed GS helps streamline workflows while ensuring patient safety during procedures.
- Integrated AV2+ Ventilator – Offers volume and pressure control ventilation modes to accommodate various patient needs and surgical requirements.
- Dual-Cylinder Yoke System – Supports oxygen and nitrous oxide cylinders, ensuring a continuous and reliable gas supply.
- Precision Vaporizers – Built-in vaporizers provide accurate agent delivery, enhancing anesthesia control for optimal patient care.
- Oxygen Failure Protection – Automatic cut-off mechanism ensures patient safety by preventing the delivery of hypoxic gas mixtures.
- Spacious Work Surface – Large tabletop and built-in storage drawers provide ample space for tools and accessories, improving efficiency.
- Compact and Mobile Design – The unit’s streamlined footprint and sturdy casters allow for easy maneuverability within the operating room.
- Breathing System Compatibility – Designed to integrate seamlessly with standard breathing circuits, ensuring operational flexibility.
Drager Narkomed GS Specifications:
Dimensions
- Height: 53.5 in (136 cm)
- Width: 36 in (91 cm)
- Depth: 27 in (69 cm)
- Weight: 375 lb (170 kg)
Gas Delivery System
- Pipeline inlet connections: DISS/male; Nut with nipple (Canada)
- Pipeline inlet pressure: 50–55 psi (345–380 kPa) (O2, N2O, Air)
- Pipeline gauge accuracy:
- ±3 psi (0–25 psi)
- ±2 psi (26–75 psi)
- ±3 psi (76–100 psi)
- Cylinder connections: Pin-indexed hanger yokes (CGA V-1-1994)
- Regulator safety relief valve: 95 psi (655 kPa)
- Regulator safety relief valve (Canada): 75 psi (520 kPa) (CSA Standard Z168.3-M84)
- Fresh gas common outlet: 15 mm female (Canada: 15 mm female, 22 mm male)
- Fresh gas oxygen concentration (ORC): 25 ±4%
- Oxygen flush flow rate: 55 (±10) l/min
- Minimum oxygen flow (at 50 psi pipeline pressure): 150 ±50 ml/min
- Low oxygen supply pressure alarm: 34–40 psi
- Cylinder gauge accuracy: ±90 psi (0–750 psi)
Vaporizers (Vapor 19.1 and 19.3)
- Temperature Range: +15–35°C (at normal atmospheric pressure of 760 mmHg)
- Flow Range: 0.25–15 l/min
- Maximum Pressure Load: 150 mmHg (above atmospheric)
- Maximum Angle of Inclination: 45°
- Weight: approximately 7.5 kg
Halothane Vaporizer
- Adjustment range: 0.2–5 vol%
- Accuracy: ±0.15% concentration (volume) or ±15% (whichever is higher)
Enflurane Vaporizer
- Adjustment range: 0.3–7 vol%
- Accuracy: ±0.2% concentration (volume) or ±20% (whichever is higher); +20%/-30% with flow settings 6.0–15 l/min and handwheel settings higher than 5.0% volume concentration
Isoflurane Vaporizer
- Adjustment range: 0.2–5 vol%
- Accuracy: ±0.15% concentration (volume) or ±15% (whichever is higher)
Sevoflurane Vaporizer
- Adjustment range: 0.3–8 vol%
- Accuracy: ±0.2% concentration (volume) or ±20% (whichever is higher)
Ventilator
- Frequency: 1–99, ±1 BPM (in 1 BPM increments)
- I:E ratio: Standard range: 1:1–1:4.5, ±0.1 (in increments of 0.5); Extended range: 4:1, 3:1, 2:1
- Inspiratory flow: 10–100 l/min (uncalibrated)
- Tidal volume: 20–1500 ml, ±100 ml
- Pressure limit control adjustment range: 15–120 cm H2O
Absorber System
- Inspiratory Valve: Mounting ring nut size: M35 x 1; Hose terminal: 22 mm male
- Expiratory Valve: Mounting ring nut size: M33 x 1; Hose terminal: 22 mm male
- PEEP Valve (optional): Range: approx. 2–15 cm H2O (continuously adjustable)
- Breathing System Pressure Gauge: Range: -20 to +80 cm H2O;Smallest scale division: 2 cm H2O; Nominal accuracy: -20 to +5 cm H2O: 3% FS; +5 to +55 cm H2O: 2% FS; +55 to +80 cm H2O: 3% FS; Mounting ring nut size: 1 1/8 x 18
- APL Valve: Nominal low flow resistance: 2 cm H2O at 8 l/min; Hose terminal: 19 mm male
- Breathing Bag Terminal: Bag terminal: 22 mm male
Oxygen Monitoring
- Range: 10–100 vol % O2
- Resolution: 1 vol % O2
- Accuracy: ±3 vol % O2 (When calibrated within 18 hours, and constant temperature and pressure)
- Response time: ≤ 25 sec (T90)
- Zero drift: ≤ 0.1 vol % O2/month
- Span drift: ≤ 1 vol % O2/8 hours
- Temperature error: ≤ ±3% of reading (15° to 40°C)
- Sensor service life: ≥ 8 months at 25°C, 50% relative humidity, 50% O2 gas mixture (or ≥5000% hour CO2)
Breathing Pressure Monitoring
- Numeric display range: -10–125 cm H2O
- Resolution: 1 cm H2O
- Accuracy: ±3 cm H2O or ±10% of reading, whichever is greater
- Waveform display range – full: 0–100 cm H2O
- Waveform resolution: 1 cm H2O
- Waveform accuracy: ±3 cm H2O or ±10% of reading, whichever is greater
- Waveform display scales: 0–20, 0–50, 0–100 cm H2O
Respiratory Volume Monitoring
- Minute Volume: Display Range: 0.1–50.0 l; Resolution: 0.1 l; Accuracy: 10% of reading or 0.01 l x breath rate, whichever is greater*
- Tidal Volume: Display Range: 0.01–2.0 l (Note: the standard bellows will deliver up to 1.5 l); Resolution: 0.01 l; Accuracy: 10% of reading or 0.015 l, whichever is greater; Volume Apnea Threshold: 0.02 l
- Respiratory Rate: Numeric display range: 2–99 bpm; Resolution: 1 bpm; Accuracy: ±10% of reading or 1 bpm, whichever is greater
Serial Interface
- Type: RS-232/422
Refurbished Drager Narkomed GS Anesthesia Machine
Soma Tech Intl offers the Drager Narkomed GS anesthesia machine up to 50% below OEM prices, providing the same service and warranty as a new unit. Each Drager anesthesia machine undergoes a comprehensive refurbishment process to ensure both optimal functionality and a like-new appearance. Our in-house certified biomedical engineers meticulously inspect, test, and replace any necessary components to ensure the Narkomed GS meets the original manufacturer’s engineering specifications. This guarantees that the anesthesia system operates reliably, just as it did when it originally left the manufacturer.
After the Drager Narkomed GS anesthesia machine passes all functional tests, it undergoes a full cosmetic restoration. The unit is thoroughly cleaned, surface imperfections are repaired, and it is repainted if necessary. New decals and labels are applied to ensure it looks as professional as it functions. Soma Tech Intl is ISO 13485 certified, ensuring that all refurbished anesthesia machines meet the highest quality standards. Before being labeled as patient-ready, the Drager Narkomed GS undergoes a final quality control inspection to verify proper operation and safety compliance.
Soma Tech Intl is actively looking to purchase used Drager Narkomed GS anesthesia machines. If your facility is looking to sell or trade in your Narkomed GS, contact Soma Tech Intl for a competitive offer. Our experienced purchasing team will guide you through the process, ensuring a seamless transaction. Selling your pre-owned Drager equipment directly to Soma removes the middleman, ensuring you receive the highest value. Soma Tech Intl is 100% HIPAA compliant, guaranteeing the secure handling of all medical equipment. If you have a Drager Narkomed GS or other anesthesia machines to sell, reach out to us today.