Covidien Valleylab Force 2 Electrosurgical Unit

December 15, 2016
Covidien Valleylab Force 2 Electrosurgical Unit

Covidien Valleylab Force 2 Features

  • High Power Digital ESU.
  • High performance capabilities in a multipurpose generator
  • Provides flexibility through varying degrees of homeostasis
  • Operator can choose one of three preset modes: simultaneous independent coagulation, low voltage coagulation and Valleylab REM safety, which substantially reduces the risk of patient burns at the return electrode site

Valleylab Force 2 Specifications

Weight

  • 23 lbs. (10.4 kg)

Dimensions

  • 8″H x 13″W x 21″L (20 x 33 x 53 cm)

Output Characteristics

  • Pure Cut: 300 watts
  • Blend 1: 250 watts
  • Blend 2: 200 watts
  • Blend 3: 150 watts
  • Coag: 120 watts
  • Low Voltage Coag: 99 watts
  • Bipolar: 70 watts

Output Waveform

  • Cut: 500 kHz sinusoid
  • Blend 1: 500 kHz bursts of sinusoid at 50% duty cycle recurring at 31 kHz
  • Blend 2: 500 kHz bursts of sinusoid at 37.5% duty cycle recurring at 31 kHz
  • Blend 3: 500 kHz bursts of sinusoid at 25% duty cycle recurring at 31 kHz
  • Coag: 500 kHz damped sinusoidal bursts with repetition frequency of 31 kHz
  • Low Voltage Coag: 500 kHz bursts of sinusoid at 25% duty cycle recurring at 31 kHz
  • Bipolar: 500 kHz sinusoid, unmodulated

Input Power Source

  • Max.Current: Idle: 0.4 A; Cut: 6 A; Coag: 3 A; Bipolar: 1.5 A
  • Max. Power: Idle: 50 W; Cut: 700 W; Coag: 300 W; Bipolar: 180 W

 