Covidien Valleylab Force 2 Electrosurgical Unit
December 15, 2016Covidien Valleylab Force 2 Features
- High Power Digital ESU.
- High performance capabilities in a multipurpose generator
- Provides flexibility through varying degrees of homeostasis
- Operator can choose one of three preset modes: simultaneous independent coagulation, low voltage coagulation and Valleylab REM safety, which substantially reduces the risk of patient burns at the return electrode site
Valleylab Force 2 Specifications
Weight
- 23 lbs. (10.4 kg)
Dimensions
- 8″H x 13″W x 21″L (20 x 33 x 53 cm)
Output Characteristics
- Pure Cut: 300 watts
- Blend 1: 250 watts
- Blend 2: 200 watts
- Blend 3: 150 watts
- Coag: 120 watts
- Low Voltage Coag: 99 watts
- Bipolar: 70 watts
Output Waveform
- Cut: 500 kHz sinusoid
- Blend 1: 500 kHz bursts of sinusoid at 50% duty cycle recurring at 31 kHz
- Blend 2: 500 kHz bursts of sinusoid at 37.5% duty cycle recurring at 31 kHz
- Blend 3: 500 kHz bursts of sinusoid at 25% duty cycle recurring at 31 kHz
- Coag: 500 kHz damped sinusoidal bursts with repetition frequency of 31 kHz
- Low Voltage Coag: 500 kHz bursts of sinusoid at 25% duty cycle recurring at 31 kHz
- Bipolar: 500 kHz sinusoid, unmodulated
Input Power Source
- Max.Current: Idle: 0.4 A; Cut: 6 A; Coag: 3 A; Bipolar: 1.5 A
- Max. Power: Idle: 50 W; Cut: 700 W; Coag: 300 W; Bipolar: 180 W