Ethicon Gen 11 / EES Harmonic Scalpel

June 12, 2017
Ethicon Gen 11 / EES Harmonic Scalpel

Harmonic scalpels use sound vibrations to cut through tissues. They are a unique version of scalpels.

Ethicon Gen 11 / EES Features

The Ethicon Endo—Surgery Generator combines technology, multiple functions, and a simple to use touch screen into one easy-to-operate unit. The technology platform supports current and future HARMONIC® and ENSEAL® devices.

  • Touchscreen for easy setup, and onscreen diagnosis
  • Universal connector and automatic instrument recognition
  • Compact design takes up less space in the OR
  • High-resolution display with wider viewing angles
  • Software updates via USB memory stick

Ethicon Gen 11 / EES Specifications

Overall Dimensions

  • Generator: 13.76” (35.0cm) x 13.98” (35.5cm) x 5.34” (13.6cm)
  • Cart: 18.88” (48.0cm) x 22.11” (56.2cm) x 37.53” (95.3cm)
  • Footswitch: 13.45” (34.2cm) x 7.50” (19.0cm) x 4.10” (10.4cm)

Weight

  • Generator: 13 lbs (5.9 kg)
  • Cart: 37 lbs (16.8 kg)
  • Footswitch: 8 lbs (3.6 kg)

Main Fuses

  • T5AL 250 V (Time-delay, 5 Amp, glass body, 5x20mm package size, quantity: 2)

Degree of protection against electric shock

  • Type CF applied part consisting of the HARMONIC® or ENSEAL® instruments

Class of protection against electric shock

  • Class 1

Ingress protection against G11 enclosure

  • 1P21

Ingress protection against G11 footswitch

  • 1P68

Main Input

  • 100-240 V, 50/60 Hz, 500 VA

Output

  • HARMONIC® Output: 150 VAC RMS maximum 35 watts continuous, 30–80 kHz (55.5 kHz unless otherwise marked in instrument IFU)
  • ENSEAL® Output: Bipolar no neutral electrode required, 100 VAC RMS maximum, 135 watts maximum, 300–490 kHZ (330 kHz unless otherwise marked in instrument IFU)

Ambient operating conditions

  • Temperature: 59°F–80.6°F, 15°C–27°C
  • Humidity: 30%–75% non-condensing
  • Atmospheric Pressure Range: 700 hPa–1060 hPa

Transport and storage conditions

  • Temperature: -31°F–129.2°F, -35°C–+54°C
  • Humidity: 10%–95% non-condensing
  • Atmospheric Pressure Range: 700 hPa–1060 hPa

 