The Zoll R Series is an ALS defibrillator, BLS defibrillator, and monitor that can provide CPR, defibrillation, and pacing therapies. The Zoll R is not just known as being a defibrillator it offers 3-lead ECG, pacing, end-tide Co2, Non-invasive Blood Pressure (NIBP), and SpO2 Monitoring. With wireless connectivity, the R Series defibrillator can give health providers a quick overall snapshot of the patient’s health via a PC or mobile device. With a color display screen, you can view the CPR dashboard which can show depth, rate, and visual indicators of compression releases. Request a quotation for a price or a certain configuration/package. Soma also offers crash carts that can be used with this series defibrillator. The previous model of the Zoll Defibrillator is the Zoll E Series.

  • Always Code-Ready
  • Remote viewing and administration from a PC or mobile device
  • History of readiness logs for the Joint Commission
  • Readiness check includes defib, pacing, and ECG modules; AC power; battery; 30-joule shock test; electrodes attached and not expired
  • Customizable dashboard

Zoll Defibrillator R Series Specifications

Dimensions

  • Height: 8.2” (20.8 cm)
  • Width: 10.5” (26.7 cm)
  • Depth: 12.5” (31.7 cm)
  • Weight: 15.2 lbs (6.9 kg)

Power

  • Voltage: 100-120 V AC
  • Frequency: 50/60 Hz
  • Battery: Rechargeable Lithium-Ion battery pack

Defibrillation

  • Waveform: Rectilinear Biphasic
  • Patient Impedance Range: 15 to 300 ohms
  • Energy Selections: 1 to 10, 15, 20, 30, 50, 75, 100, 120, 150, and 200 joules selected using controls on the front of the defibrillator or sternum paddles. (Note: When using appropriate pediatric resuscitation electrodes, the 75-joule setting is replaced by 70- and 85-joule settings.)
  • Smart Step Energy Levels: Automatically escalates energy through a configured adult or pediatric protocol
  • Energy Display: Shown on the monitor for both selected and delivered
  • Charge Time: Less than 7 seconds with a new, fully charged battery (first 15 charges to 200 joules); longer charge times may result in a depleted or older battery
  • Synchronized Mode: Synchronizes defibrillator pulse to patient’s R wave. “SYNC” message displayed on monitor and markers shown on both monitor and recorded ECG
  • Charge Controls: Control from the front of the defibrillator or apex paddle
  • Paddles: External apex/sternum paddles; adult plates slide off to expose pediatric electrode surface
  • Code Readiness Testing (OneStep): Verifies defibrillator hardware, therapy delivery cable (with both paddles and electrodes), electrode condition, and expiration (with select OneStep electrodes) without the need for a separate test fixture.

ECG Monitoring

  • Patients Connection: 3-lead ECG cable, 5-lead ECG cable, paddles, or hands-free electrodes; selectable by the front panel switch
  • Input Protection: Fully defibrillator protected. Circuits designed to prevent distortion of ECG signal by pacer pulse
  • Implanted Pacemaker Spike Display: Circuits designed to detect most implanted pacemaker spikes and display a marker on the ECG trace
  • Bandwidth: 0.5 to 21 Hz (-3dB) standard; 0.05 to 150 Hz diagnostic with configurable options of 0.5 Hz to 40 Hz or 1 Hz to 21 Hz
  • Lead Selection: I, II, III aVR, aVL, aVF, V, P1, P2, P3 with OneStep Pacing electrode
  • ECG Size: 0.5, 1.0, 1.5, 2.0, or 3.0 cm/mV display on the monitor
  • Heart Rate Display: 0 to 300 bpm ±5%
  • Heart Rate Alarm: User selectable for tachycardia at 60 to 280 bpm; for bradycardia at 20 to 100 bpm. On/Off status displayed on the screen.

Display

  • Screen Type: Color, VGA liquid crystal display (LCD)
  • Screen Size: 6.5 inches (16.5 cm) diagonally
  • Sweep Speed: 25 mm/sec
  • Viewing Time: 5 seconds with the standard display format
  • Channels: 3
  • Information: Heart Rate, Leads/Pads, Alarm On/Off, Selected Energy, Delivered Energy, User Prompts and Warnings, Code Readiness Test Results, SpO2, Non-invasive Blood Pressure, EtCO2, Pacer Functions, Code Markers, CPR Dashboard

Battery Packs

  • Type: 10.8 V (nominal) rechargeable lithium-ion
  • Capacity: 5.8 amp hours
  • Weight: 1.7 lb (0.77 kg)
  • Recharge Time: 5 hours or less with an integral charger
  • Operating Time: >4 hours of continuous ECG monitoring; 100 maximal energy (200 joules) discharges; 3.5 hours of continuous ECG monitoring and pacing at 60 mA, 80 ppm

Recorder

  • Technology: 90 mm thermal array; 80 mm grid width
  • Speed: 25 mm/sec, 6-second delay
  • Zoll Printing Modes: Manual or automatic
  • Annotations: Time, date, defibrillation energy, patient impedance, heart rate, pacer output, QRS synchronization marker, ECG size, ECG lead, alarm, defibrillator test results, analyze ECG, ECG bandwidth

External Pacing

  • Type: VVI demand; asynchronous (fixed rate) when used without ECG leads or in asynchronous (ASYNC) pacing mode
  • Pulse: Rectilinear, constant current: 40 ms ±2 ms; variable 0 to 140 mA ±5% or 5 mA, whichever is greater. The rate is variable from 30 to 180 ppm ±1.5%
  • Output Protection: Fully defibrillator protected and isolated
  • Zoll Pacing: Eliminates the need to connect separate ECG leads and or a cable when used in conjunction with official electrodes.

Pulse Oximetry with Masimo SET Technology

  • Saturation Range: 1-100% (%SpO2 ) with a resolution of 1%
  • Pulse Rate Range: 25-240 ppm with a resolution of 1 ppm
  • Saturation Accuracy: Non-motion conditions ±2% for adults/pediatrics; ±3% for neonates. During motion ±3% for all patients
  • Pulse Rate Accuracy: Non-motion conditions ±3 ppm. During motion ±5 ppm.

Mainstream CO2 Capnostat 5 Sensor

  • Principle of Operation: Nondispersive infrared (NDIR) single-beam optics, dual-wavelength, no moving parts
  • Warm-up Time: Full specifications within 2 minutes at an ambient temperature of 25°C. Capnogram in 20 seconds
  • Environmental: Zoll Operating Temperature: 0°C to 45°C, Storage and Shipping Temperature: -40°C to 70°C.

Sidestream CO2 LoFlo Sensor

  • R Series Principle of Operation: Nondispersive infrared (NDIR) single-beam optics, dual-wavelength, no moving parts
  • R Series Warm-Up Time: Full specifications within 2 minutes at an ambient temperature of 25ºC. Capnogram in 20 seconds
  • R Series Environmental: Operating Temperature: 0°C to 40°C, Storage and Shipping Temperature: -40°C to 70°C.

NIBP

  • Zoll R Series Patients Type: Adult, Pediatric
  • Zoll R Series Method: Oscillometric
  • Zoll R Series Control: Automatic and manual non-invasive blood pressure measurement.

WiFi Capable

  • WiFi 802.11 a/b/g/n Ambicom-specific 1100C-CF Card P/N 8005-000101-01 compatibility
  • Typical Readiness File Data Size: 750K
  • Zoll Code Data File: 1.2 MB.

Accessories

  • All equipment accessories can be found on the Soma medical equipment accessories and parts website.
  • 8009-0762-01 – OneStep Leads – AHA
  • 9650-0903-01 – Service Manual
  • 8900-000219-01 – OneStep Pediatric CPR Electrode 1ct
  • 8900-000220-01 – OneStep Pediatric CPR Electrode 8ct
  • 8900-0214-01 – Resuscitation Electrode 8ct
  • 8009-0750 – Pacing Cable
  • 8900-0213-01 – CPR Resuscitation Electrode 8ct
  • 8900-0211-01 – Resuscitation Electrode 8ct