Medtronic Covidien Barrx Flex Energy Generator
24-Hour QuotesPricing up to 50% below the cost of new,
with service and warranty included.
Medtronic Covidien Barrx Flex Energy Generator Features
The Medtronic Covidien Barrx Flex Energy Generator is a bipolar radiofrequency (RF) energy generator that is designed to precisely control depth and uniformity of tissue ablation in the GI tract. The Barrx Flex provides you with the flexibility to use the Barrx 360 RFA balloon catheter for larger circumferential treatment areas or one of the Barrx family of focal catheters for the treatment of smaller focal areas. This RF generator automatically recommends ablation catheter sizing, tracks, and displays the total number of ablations, and automatically adjusts default power density settings to match the catheter. Optional accessories include the Flex RFA footswitch and the Flex RFA output cable.
- Seamless transition between circumferential and spot treatment cases.
- Intuitive sizing provides improved differentiation between the esophagus and stomach.
- Rapid firing for reduced procedure time.
- Recovery in <2 seconds with Barrx RFA ablation catheters.
- Automatically recommends ablation catheter sizing.
- Tracks and displays the total number of ablations.
- Automatically adjusts default power density settings to match the catheter.
- A platform for future clinical and product innovation.
Medtronic Covidien Barrx Flex Energy Generator Specifications
Dimensions
- Height: 6.13″ (15.9 cm)
- Width: 16″ (39.4 cm)
- Depth: 21″ (51.6 cm)
- Weight: 31 lb. (14 kg)
Measurement Accuracy
- Power: ±15%, ±3 W over the resistance load range of 0.5 O to 20 O for power levels above 30 W
Environmental Specifications
- Operational temperature: 20° C to 30°C
- Humidity functional: 20% to 65% non-condensing
- Storage temperature: -40° C to 70° C
- Humidity storage: 20% to 85% non-condensing
- Altitude: up to 7000 feet (2134 meters) above sea level
- Barometric pressure: Sea Level (1013.25 hPa) to 788 hPa (7000 ft.)
Fuses and Line Voltage Ranges
- Line Voltages: 110 – 120 VAC/60 Hz and 220 – 240 VAC/50 Hz
- Rated input current: <6.3 A, 110-120 VAC, 60 Hz, or <4.0 A, 220-240 VAC, 50 Hz
- Replace mains fuses as follows:
- Littelfuse p/n 021506.3 (6.3 A) for the 110–120 VAC/60 Hz System
- Littelfuse p/n 0215004P (4.0 A) for the 220–240 VAC/50 Hz System
- Power Output Restrictions: Power output is available over a 0.4 O – 35 O range