Vyaire 3100a

24-Hour Quotes

Pricing up to 50% below the cost of new,
with service and warranty included.
Download Specifications

Vyaire 3100a Ventilator Features

The Vyaire 3100a is a High Frequency Oscillatory Ventilator. This kind of ventilation is intended to help treat respiratory failure in neonatal patients and acute respiratory distress in pediatric patients. With tools that can protect the lungs from damage, the Vyaire 3100A can expend pressure and volume to inflate the lungs and prevent a neonatal patient from going into respiratory failure. The 3100A can help ensure the patient’s care while providing the required ventilation.

  • Gently delivers 1 to 3 mL tidal volumes to ventilate the most premature infant and delivers up to 180 mL to support the ventilation of pediatric patients.
  • Produces an active exhalation,4 which is essential at high frequency respiratory rates to prevent air trapping that may occur with passive exhalation.
  • Offers patented technology and a highly reliable, electromagnetically driven piston that distinguishes it from other high frequency ventilators.
  • Permits variable I:E ratios, which are desirable for managing ventilation and reducing the risk of air trapping.

Vyaire 3100a Ventilator Specifications

Dimensions

  • Height: 53.8 In (136.7 cm)
  • Width: 18.6 In (47.2 cm)
  • Depth: 11.4 In (29 cm)
  • Weight: 143 lbs (64.9 kg)

Bias Flow

  • 0–40 liters per minute (LPM) Continuous, 15-turn control.
  • Resolution: 2.5 LPM.
  • Accuracy: ±10% of full scale at the following conditions: air or oxygen @ 70°F and 760 Torr.

Mean Pressure Adjust

  • Approximately 3–45 cmH2O minimum range; Bias Flow dependent.
  • Resolution: 0.1 cmH2O on airway pressure digital meter, 1-turn control.
  • Accuracy: Non-calibrated control knob.

Mean Pressure Limit

  • Approximately 10–45 cmH2O mean proximal airway pressure.
  • Resolution: 0.1 cmH2O on airway pressure digital meter, 1-turn control.
  • Accuracy: Non-calibrated control knob.

Power

  • At 100% power, ?P >90 cmH20 max amplitude of proximal airway pressure.
  • Resolution: Graduated 10-turn locking dial, not calibrated in % power.

Frequency-Hz

  • 3–15 Hz oscillator frequency.
  • Resolution: 0.1 Hz on digital meter, 10-turn control.
  • Accuracy: ±5% of full scale.

% Inspiratory Time

  • 30–50% of oscillatory cycle.
  • Resolution: +/- 1% as read on digital meter
  • Accuracy: ±5% of full scale.

Set Max Paw Alarm Thumbwheel

  • 0–49 cmH2O mean airway pressure.
  • Resolution: 1 cmH2O
  • Accuracy: Within ±2 cmH20.

Set Min Paw Alarm Thumbwheel

  • 0–49 cmH2O mean airway pressure.
  • Resolution: 1 cmH2O.
  • Accuracy: Within ±2 cmH20.
Banner

Questions about our ventilators?

Soma Tech Intl offers a wide range of ventilators. If you have any questions about any of our ventilators or need a quick quote, call us and one of our knowledgeable sales representatives will help you.

Looking to sell your ventilators? Soma ensures your facility gets top dollar for your equipment and also gives the option for trade-in credit.

1-800-GET-SOMA
Contact Us