GE Marquette Mac 5000 Features
The GE Marquette MAC 5000 is an ECG system primarily used for resting ECG analysis readings. MAC 5000
can provide advanced ECG readings for both a 12 and 15-lead analysis, interpretation, and waveforms. The EKG system comes with, a 10-inch display, wireless communications, and networking capabilities. The GE System weighs around 15 pounds making it easy to transport by itself or by a cart to the patient.
- 15 and 12 lead EKG analysis
- Offers stress testing and resting ECG
- Gender-specific criteria for identifying arrhythmia
- 10-inch color display
- Adult and pediatric interpretation
GE Marquette Mac 5000 Specifications
Dimensions (without trolley)
- Height: 3.7 in (9.4 cm) (with display closed)
- Width: 15 in (38.1 cm)
- Depth: 13.8 in (35.1 cm)
- Weight: 15 lb (6.8kg)
Operational Conditions
- Temperature: 10°C to 40°C (50°F to 104° F)
- Relative Humidity: 20% to 95% RH noncondensing
- Atmosphere Pressure: 700 to 1060 hPa
Processing
ECG Interpretation
- 12SL Adult and Pediatric ECG Analysis Program
Computerized Measurements
- 12SL™ Adult and Pediatric ECG Analysis Program
Optional
- PHI-RES & HI-RES Late Potential Analysis
Additional ECG Function
- Vectorcardiography
ECG Analysis Frequency
- 500 samples/second (sps)
ECG Storage
- 150 typical
- 200 ECGs maximum on removable media (1.44 MB, 3.5diskette)
Digital Sampling Rate
- 4000 samples / second / channel
Pre-Acquisition
- Provides 10 seconds of instantaneous ECG acquisition
AC Differential
- ±10 mV DC offset: ±320 mV
Resolution
- 4.88 µV/LSB @ 500 sps
Frequency Response
- –3 dB @ 0.01 to 150 Hz
Common Mode Rejection
- >140 dB (123 dB with AC filter disabled)
Input Impedance
- >10M@ 10 Hz, defibrillator protected
Patient Leakage
- <10 µA
Pace Detect
- Orthogonal LA, LL and V6; 750 µV @ 50 µs
Special Acquisition Functions
- Disconnected lead detection
- Electrode impedance
- Excessive AC noise
- Baseline wander and muscle tremor messages
Heart Rate Meter
- 30 to 300 BPM ± 10% or 5 BPM whichever is greater. Heart rates outside this range will not be displayed
Display
Display Type
- 264 mm (10.4 in) diagonal graphics backlit AM LCD
Display Resolution
- 640 x 480 pixels with waveform enhancement
Display Data
- Heart rate
- Patient name
- ID
- Clock
- Waveforms
- Lead labels
- Speed
- Gain and filter settings
- Warning messages
- Prompts and help messages
Writer
Writer Technology
- Thermal dot array
Writer Speeds
- 5 mm/s
- 12.5 mm/s
- 25 mm/s
- 50 mm/s (same as display)
Number of Traces
- 3
- 6
- 12
- 15
- user selectable (same as display)
Writer Sensitivity / Gain
- 2.5, 5, 10, 20, 10/5 (split calibration) mm/mV (same as displayed)
Writer Speed Accuracy
- ±2%
Writer Amplitude Accuracy
- ±5%
Writer Resolution
- Horizontal 1000 dpi @ 25 mm/s
- 200 dpi vertical
Paper Type
- Thermal, Z-fold, perforated, fan fold, 300 sheets/pack
Paper Size
- A size: 214.63 mm x 280 mm (8.45 in x 11 in)
- A4 size: 210 mm x 297.5 mm (8.27 in x 11.7 in)