Datascope Spectrum

Quotes in less than 24 hours Download Specifications

Datascope Spectrum OR Features

The Datascope Spectrum OR is a patient monitor that can be used in an operating room setting. The Spectrum offers readings for additional measurements like ECG, SpO2, NIBP, Spirometry, temperature, and BISx. The Datascope system is based on the same platform as the Passport 2, offering either a Masimo or Nellcor SpO2 within the system. The system is easy-to-use with a 12-inch screen display monitor, quick-access menus, and one-touch functionality.

  • 12-inch high- contract color display
  • Auto-adjustable display for optimal visibility
  • 3-8 waveform configurations
  • BIS Consciousness monitoring
  • Includes Pre-anesthesia monitoring features:
    • 3 or 5 lead ECG
    • Non-invasive blood pressure
    • Comes with either a Nellcor or Masimo SpO2 monitoring system

Datascope Spectrum OR Specifications

Dimensions

  • Height with Module: 10.7 in (27.2 cm)
  • WidthWith Module: 11.9 in (30.2 cm)
  • Depth with module: 7.5 in (18.8 cm)
  • Weight with modules: 1.75 lb ( 0.8kg)
  • Weight with battery: 13.52 lb (6.1kg)

Display

  • Size: 12.1-inch (30.7cm) Color Active Matrix TFT LCD
  • Resolution: 800 x 600 pixels
  • Waveforms: 3 to 8

ECG (3-lead and 5-lead)

  • Leads: I, II, III, aVR, aVL, aVF, V
  • Cable Detection: Autodetecting Datascope 3 or 5 wire
  • Display Sensitivity: 0.25, 0.5, 1.0, 2.0, 3.0, 4.0, cm/mV ± 10%
  • Frequency Response To Screen
  • Extended Mode: 0.05 – 100 Hz, -3db
  • Monitor Mode: 0.5 – 40 Hz, -3db
  • ST Mode: 0.05 – 40 Hz, -3db
  • ECG Sync Pulse for Cardioversion:
  • Delay: ≤35 ms max between QRS Peak and the rising edge of Sync Pulse
  • Amplitude: 2Vp minimum into a 5k ohm load
  • Width: 2-7 ms
  • Analog Output: (ECG)
  • Delay: 25 ms max
  • Sensitivity: 1 V/mV of input, ± 10%
  • Defibrillator Overload Protection:
  • Withstand 360 Joule discharge as per IEC 60601-2-27, 51.101.1
  • Recovery Time: Time for recovery to within 1mV in < 8 seconds automatically. 3 seconds from 1Vpp at 60 Hz

Heart Rate Meter

  • Range: 30 – 300 BPM Adult/Pediatric, 30 – 350 BPM
  • Neonate
  • Accuracy: ± 3 BPM or ± 3% at 30 – 250 BPM whichever is greater, ± 5% at 251 – 350 BPM
  • Pacer Rejection: Rejects all pulses of amplitude ± 2.0 mV to ± 700 mV and duration 0.1 to 2 ms with no tail. AAMI EC-13-1992 4.1.4
  • (3-Lead and 5-Lead) Rejects all pulses of amplitude ± 2.0 mV to 700 mV and duration 0.1 to 2ms with 100 ms time the constant tail of less than 2.0 mV, or 4 ms time the constant tail of < 2.0 mVper AAMI EC-13-1992 3.1.4.2 Tall T-Wave Rejection: Rejects all T-Waves of amplitude less than 120% of 1 mV, 100 ms QRS, T wave duration of 180 ms and Q-T interval of 350 ms per AAMI EC-13-1992 4.1.2.1 (c)

ST Analysis

  • 3-lead or 5-lead Range: -9.9 mm to +9.9 mm
  • Resolution: 0.1 mm
  • Default ST Measuring Point: 80 ms after j point for HR <120 BPM, 60 ms after j point for HR >120 BPM

Arrhythmia Analysis

  1. 3-lead or 5-lead cables: Adult / Pediatric Only: Asystole, Irregular Heart Rate, Couplets, Bigeminy, Trigeminy, Ventricular Tachycardia, Ventricular Fibrillation, PVC ’s per minute, Runs,
  2. Ventricular Rhythm

Respiration (ECG)

  • Range: 4 to 199 BPM
  • Accuracy: ± 2% or 2 breaths per minute whichever is greater from 4 to 150 BPM, ± 4% from 151 to 199 BPM
  • Lead: I or II

Non-Invasive Blood Pressure

  • Technique: Oscillometric
  • Systolic Range: Adult: 55 – 235 mmHg, Pediatric: 55 – 160 mmHg, Neonate: 45 – 120 mmHg
  • Diastolic Range: Adult: 30 – 200 mmHg, Pediatric: 30 – 150 mmHg, Neonate: 20 – 100 mmHg
  • Systolic Accuracy: Mean Error less than ± 5 mmHg, standard deviation less than ± 8 mmHg
  • Diastolic Accuracy: Mean Error less than ± 5 mmHg, standard deviation less than ± 8 mmHg
  • Pulse Rate Range: Adult/Pediatric: 35 to 245 BPM Neonate: 70 to 245 BPM
  • Pulse Rate Accuracy: ±3 BPM or 3% whichever is greater
  • Connector Type: Rectus
  • Cuff Inflation: Volume of 500 ccs to 300 mmHg in ≤35 sec

Temperature

  • Scale: Selectable C° or F°
  • Range: 15° to 45°C / 59° to 113°F (T1 and T2) 0°C to 5.5°C / 0°F to 9.9°F (Delta T)
  • Accuracy: ± 0.1°C (15°C to 45°C) exclusive of probe errors. ± 0.2°F (59°F to 113°F) exclusive of probe errors.

Pulse Oximetery

  • Masimo SET SpO2 Accuracy Saturation with no motion conditions
  • Adult/Pediatric: 70% to 100% ± 2 digits SpO2, 0 to 69% unspecified
  • Neonate: 70% to 100% ± 3 digits SpO2, 0 to 69% unspecified
  • Ear Sensor (Adult/Pediatric): 70% to 100% ± 4 digits SpO2, 0 to 69% unspecified
  • Masimo SET SpO2 Accuracy Saturation during motion conditions
  • Adult/Pediatric/Neonate: 70% to 100% ± 3 digits SpO2
  • Response Time: 18 seconds to 95% of the final step of % SpO2 value from 60-95% at 75 BPM. Averaging set at 8 seconds.
  • Pulse Rate Range Masimo with no motion conditions
  • Adult/Pediatric/Neonate: 30-235 ±3 BPM, Ear Sensor (Adult/Pediatric) 30-235 ± 3 BPM
  • Pulse Rate Range Masimo during motion conditions Adult/Pediatric/Neonate: 30-235 ±5 BPM
  • Low Perfusion Performance
  • Masimo: > 0.02% Pulse Amplitude and % Transmission > 5% / Saturation (%SpO2) ± 2 digits; Pulse ± 3 digits
  • Nellcor OxiMax SpO2 Saturation Accuracy
  • Adult/Pediatric/Neonate: 70% to 100% ± 3 digits
  • Pulse Rate Range Nellcor: 20-249 ±3 BPM

IBP

  • Range: Sys/Dia/Mean -30 to +300 mmHg
  • Accuracy: ± 2 mmHg or 2% whichever is greater
  • Scale: -10 to 10, 0-20, 0-40, 0-160, 0-225, 0-320, 60-140 mmHg
  • Zero Range: ± 120 mmHg
  • Excitation: 5V DC ± 2%
  • Frequency Response: DC to 16 Hz ± 1Hz, -3db

Recorder

  • Speed: 3.125, 6.25, 12.5, 25 mm, and 50 mm/sec
  • Note: 3.125 speed is for CO2/Resp only.

CO2 (Microstream)

  • Range: 0 – 99 mmHg
  • Accuracy: 0 – 38 mmHg ± 4 mmHg 39 – 99 mmHg ± 12% (0 – 20 min)
  • 0 – 38 mmHg ± 2 mmHg 39 – 99 mmHg ± 5% + .08% for every 1mmHg above
  • 40 mmHg (> 20 min)
  • Respiration Rate: 0-150 breaths per minute
  • Respiration Accuracy: ± 1 breath per minute from 0 – 40 BPM
  • ± 2 BPM from 41– 70 BPM
  • ± 3% from 71 – 100 BPM
  • ± 5% from 101 – 150 BPM
  • CO2 Waveform Recognition: 5 mmHg ± 2 mmHg bandwidth
  • Start-up Time: 30 seconds typical
  • Sampling Rate: 50 ml/min ± 7.5 ml/min
  • Auto Zero: <15 seconds at 5, 10, 15, and 60 minutes, then every 60 minutes thereafter.

Electrical Rating

  • AC Voltage: 100 – 240 VAC ± 10%, 50/60 Hz ± 3 Hz
  • Battery Type: Sealed Lead Acid
  • Number of Batteries: 2
  • Battery Voltage: 12 V
  • Battery Capacity: 2.3 Ah
  • Battery Run Time: 1 hour, 20 min from two fully charged, new batteries at 25° C with ECG, SpO2, and NIBP running at 15-minute intervals
  • Recharge Time: 16 hrs max

Cardiac Output

  • Range: 0.2 to 20.0 liters/min
  • C.O. Repeatability: ± 2% or 0.02 l/min from the mean value
  • Blood Temp Range: 17.5°C to 43°C (63.5°F to 109.4°F)
  • Blood Temp Accuracy: ± 0.2°C (± 0.4°F) exclusive of probe errors
  • Injectate Temp Range: -1.0°C to 30.0°C (30.2°F to 86°F )
  • Injectate Temp Accuracy: ± 0.2°C (± 0.4°F) exclusive of probe errors

BIS (Bispectral Index)

  • Bispectral Index (BIS) Range: 0 – 99
  • Signal Quality Index (SQI) Range: 0 – 100%
  • EMG Range: 30 – 80dB
  • Suppression Ratio (SR): 0 – 100%
  • Burst Count: 0 – 30
  • Input Impedance: >50 Mohm
  • Noise (RTI): <0.3uV RMS 0.25 – 50 Hz
  • Input Range: +/- 1mV
  • EEG Bandwidth: 0.25 – 100Hz (-3dB)
  • FILTERS (ON)
  • High Pass: 2.0 Hz (-3dB)
  • Low Pass: 70 Hz (-3dB)
  • Notch: 50 Hz and 60 Hz
  • FILTERS (OFF)
  • High Pass: 0.25 Hz (-3dB)
  • Low Pass: None
  • Notch: None

MAC

  • The Spectrum OR shall compute the uncorrected MAC value for all agents supported by the Gas Module SE using the following formula: Where AA is the anesthetic agent in use, ET AA is the end-tidal agent concentration, x(AA) is a clinically derived coefficient based on anesthetic
  • agent (known as 1MAC values), ET N2O is the end-tidal N2O concentration and x(N2O) is a clinically-derived coefficient for N2O (also known as the 1MAC value).
  • X(Isoflurane) = 1.15%
  • X(Halothane) = 0.77%
  • X(Sevoflurane) = 2.1%
  • X(Desflurane) = 7.3%
  • X(Enflurane) = 1.7%
  • X(N2O) = 105%
  • The uncorrected MAC value is not corrected for ambient pressure (altitude & barometric effects), patient age, core temperature, or other factors influencing volatile anesthetic agents’ effect. Reference Source: EN ISO 21647:2004 Medical electrical equipment-
  • Particular requirements for the basic safety and essential performance of respiratory gas monitors.

Refurbished of The Datascope Spectrum Patient Monitor

Soma Tech Intl offers the Datascope Spectrum Patient Monitor

up to 50% below OEM prices with the same service and warranty as new. The Datascope is not only technically refurbished but also cosmetically refurbished to make it work and look like new. When the refurbishing process starts on the monitor, it is carefully inspected and tested by our in-house biomedical engineers who are highly skilled and certified to work on patient monitor systems. If necessary, parts are replaced to make sure that all aspects of the Spectrum are to the original engineering manufacturer’s specifications. This ensures that the unit is working the same way that it originally left the manufacturer. After the patient monitor is working like new it undergoes a unique cosmetic restoration process. The monitor is cleaned, minor scratches and dents are repaired, it’s painted, and new decals are replaced if necessary. After this strenuous refurbishing process, the unit functions and looks like new. Soma Tech Intl is ISO 13485 certified and ensures that the best quality is given to all capital medical equipment they sell. Before the Datascope Spectrum is labeled patient-ready and packed up for shipment, it goes through one last quality control test that makes sure all aspects are again working as they should. If you are looking to buy a refurbished patient monitor contact soma at [email protected] or call us at 1-800-GET-SOMA.

Soma Tech Intl is actively looking to purchase a Datascope patient monitor, If you or your facility is looking to sell your patient monitor, be sure to contact Soma Tech Intl. Soma has a dedicated and experienced purchasing department that helps you through the entire process. Selling your used or pre-owned patient monitors takes out the middleman in the equation. This also ensures that your medical facility gets top dollar for your equipment. Soma is 100% HIPPA compliant and ensures that your equipment gets to our facility in the same condition that it left yours. Soma Tech Intl also gives the option for trade-in credit on your medical equipment. If you or your facility is looking to regain space for new equipment, or if you are looking to upgrade your current equipment, contact Soma Tech Intl’s purchasing department at [email protected] or send us a message on our Sell to Soma website.

Questions about our Patient Monitors?

Soma Tech Intl offers a wide range of

Patient Monitors. If you have any questions about any of our Patient Monitors or need a quick quote, call 1-800-GET-SOMA, and one of our

knowledgeable sales representatives will help you.