Mindray Passport 12

Quotes in less than 24 hours Download Specifications

Mindray Passport 12 Features

The Mindray Passport 12 is a neonatal, pediatric, and adult patient monitor used to display, review, and store ECG, arrhythmia detection, QT analysis, heart rate, respiration, temperature, SpO2, Pulse rate, non-invasive and invasive blood pressure, cardiac output, CO2, and anesthetic gas. The Passport 12 weighs in at 10 pounds, has a 12-inch high-resolution color touchscreen display, and can display up to 8 waveforms.

  • 12-inch high-resolution color touchscreen display.
  • Displays up to 8 waveforms.
  • Standard parameters include 3- or 5-lead ECG, Masimo SET, SpO2, 2 IBP, NIBP, RESP, TEMP, and an integrated recorder.
  • Features extensive data storage for trend data, alarms, events, and NIBP measurements for up to 48 hours.
  • Connects to BeneVision Central Station.
  • Integrates with eGateway for communication to EMR.
  • Defibrillator synchronization.
  • Convenient, ergonomic carrying handle and multiple mounting options.
  • Optional — Nellcor OxiMax SpO2.
  • Optional — ST and Arrhythmia analysis with atrial fibrillation detection available.
  • Optional — Wireless 2.4 GHz (802.11b/g/n) available.

Mindray Passport 12 Specifications

Dimensions

  • Height: 10.9 (27.6 cm)
  • Depth: 5.4 in (13.7 cm)
  • Width:12.7 in (32.2 cm)
  • Weight: 4.5 kg (10 lbs) (Standard parameters configuration, including 2 lithium batteries and a recorder)

Input/Output Communications ECG Analog Output

  • Bandwidth (-3dB; reference frequency: 10Hz)
  • Diagnostic Mode: 0.05 to 150 Hz
  • Monitor Mode: 0.5 to 40 Hz
  • Surgical Mode: 1 to 20 Hz
  • ST Mode: 0.05 to 40 Hz
  • Maximum Transmission Delay: =30 ms
  • Sensitivity: 1V/mV ±5%
  • Pacer Enhancement
  • Amplitude: Voh=2.5V
  • Width: 10ms±5%
  • Signal Rising and Falling Time: =100µs

Electrical Ratings

  • Power Supply: AC power, battery
  • AC Power: 100 to 240 VAC (±10%)
  • Current: 1.3 to 0.5 A
  • AC Power Input
  • Frequency: 50/60 Hz (±3Hz)

Environmental Conditions

  • Temperature Range Operating: 10°C to 40°C
  • Temperature Range Non-operating (storage): -20°C to 60°C
  • Humidity Range Operating: 15% to 95%, non-condensing
  • Humidity Range Non-operating (storage): 10% to 95%, non-condensing
  • Altitude Range Operating: -1,000 to 9,841 feet ASL)
    • 1,074 hPa to 700 hPa
  • (806 mmHg to 394 mmHg)

Battery

  • Type: Chargeable Lithium-Ion, 11.1DVC, 4.5 Ah
  • Number of Batteries: 2
  • Run Time: 4 hours when powered by one new fully-charged battery (25°C, SpO2, ECG, disconnected from Temp cable, Auto NIBP measurements at intervals of 15 minutes)
  • 8 hours when powered by two new fully-charged batteries (25°C, SpO2 sensor, and ECG cable connected, Temp cable not connected Auto NIBP measurements at an interval of 15 minutes)
  • Charge Time: Less than 3 hours to 90%, and less than 4 hours to 100% when the monitor is off.
  • Less than 8 hours to 90%, and less than 12 hours to 100% when the monitor is on.
Banner

Questions about our Patient Monitors?

Soma Tech Intl offers a wide range of patient monitors. If you have any questions about any of our patient monitors or need a quick quote, call us and one of our knowledgeable sales representatives will help you.

Looking to sell your patient monitors? Soma ensures your facility gets top dollar for your equipment and also gives the option for trade-in credit.

1-800-GET-SOMA
Contact Us