Baxter InfusOR Infusion Pump
November 10, 2016Baxter InfusOR Features:
• Designed to aid in many intravenous agents given during anesthetic procedures.
• Provides for the convenient delivery of narcotics, muscle relaxants, vasoactives and other drugs routinely given during anesthesia.
• Easy to use and set up.
• Drug delivery choices are easily selected and changed using front panel rotary switches.
Baxter InfusOR Specifications:
Accuracy:
• Infusion: linear rate ± 3%. Bolus: linear displacement ± 3%.
Battery Life:
• Typical use: 150 hours. At low battery and typical use: 4 hours
Batteries:
• Four “C” – size Alkaline Cells.
Operating Voltage:
• 5 – 7 Volts.
Operating Current:
• 20 mA.
Low Battery Voltage:
• 5.0 Volts approximately.
Flow Rates:
• 0 to 600 mL/hr.
Flow Profile:
• Bolus: continuous.
• Infusion: pulsed intermittently.
Occlusion Force:
• 8 ± 1 lbs.
Maximum Occlusion Pressure:
• 11 psi w/60 cc syringe.
• 20 psi w/20 cc syringe.
Occlusion Detection Time:
• 120 seconds at 36 mL/hr. (inverse related to flow rate)
Volume Stored on Occlusion:
• 1.1 mL approximately.
Effect on Accuracy:
• None to Occlusion Pressure.
Size:
• 9.2 x 4.5 x 2.0 inches.
Weight:
• 2 lbs. With batteries.
Syringes:
• Monoject® or B-D® 60 cc.
• Monoject® or B-D® 20 cc.