Philips BV Pulsera
November 30, 2016Philips BV Pulsera Features
The Philips BV Pulsera C-Arm is a multi-application mobile X-ray imaging system for the most challenging interventional procedures in Pain Management, General, Orthopedic, Vascular, Neurovascular and Cardiac surgeries. The Philips BV Pulsera is a powerful fluoroscopy system with a high resolution CCD camera, outstanding image quality with low X-ray dosage, crisp, clear images and superior penetration power.
Philips BV Pulsera Specifications
X-Ray System
Specialty Packages Available
- Extended processing.
- Vascular functions.
- Extended vascular functions.
- Cardiac Pulsed functions.
Generator
- 80 kHz High frequency.
- 7.5 kW full-wave.
- Up to 120 kVp.
- Up to 60 mA for radiographic film exposures.
- Continuous fluoro boost mode up to 20 mA.
- Pulsed fluoro mode.
- Pulsed fluorography mode up to 60 mA.
- Full power from outlet:
- 50 Hz or 60 Hz.
- 100V-20A to 240V-10A.
X-Ray Tube
- Rotating anode X-ray tube.
- Focal spots: 0.3 mm and 0.6 mm.
- Anode heat capacity: 300,000 HU.
- Anode cooling rate: 70,000 HU/min.
- Housing heat capacity: 1,900,000 HU.
- Anode diameter: 2.8 in.
- Anode angle: 10°.
Collimator
- Iris collimation with dual opposing lead shutters allowing elliptical, round and rectangular collimation.
- Minimum beam size < 5 X 5 cm2.
- Iris and shutters are continuously variable.
- All functions remotely controlled from the C-Arm control panel.
Fluoroscopy Mode
- Focal spot: 0.3 mm and 0.6 mm.
- kVp range: 40 – 120 kVp.
- mA range: 0.2 – 20 mA continuous High Definition fluoro boost mode.
- Auto and manual modes.
- Continuous or Pulsed operation.
- APF varies mA, kVp and camera gain.
- User specific APF tables.
Pulsed Fluoroscopy Mode
- Half Dose Fluoroscopy.
- Quarter Dose Fluoroscopy.
- Focal spot: 0.6 mm.
- kVp range: 40 – 120 kVp.
- mA range: 0.4 – 12 mA.
- Pulse rate: 6.25 or 12.5 pulses per second.
- Pulse width: 11 to 37 milliseconds.
- Computer controlled APF, mA, kVp and camera gain.
Pulsed Fluorography Mode
- Focal spot: 0.6 mm.
- kVp range: 40 – 110 kVp.
- mA range: Up to 60 mA.
- Pulse rate: 3 to 30 pulses per second. (optional)
- Pulse width: 8.0 / 9.5 / 11.1 milliseconds.
- Computer controlled camera iris, mA, kVp, & camera gain.
Digital Exposure Mode
- Focal spot: 0.6 mm.
- kVp range: 40 – 110 kVp.
- mA range: 1.0 – 75 mA.
Radiographic Mode
- Focal spot: 0.6 mm.
- kVp range: 40 – 120 kVp.
- Fixed 60 mA.
- 0.6 mm – mAs range: 3.2 – 125 mAs.
Video Imaging System
Image Intensifier
- 9 in Image Intensifier
- Tri-mode 9 in/7 in/5 in image intensifier
- 9 in.: 1.6 lp/mm
- 7 in.: 2.1 lp/mm
- 5 in.: 2.9 lp/mm
- DQE: 65% (typical)
- Tri-mode 9 in/7 in/5 in image intensifier
- 12 in Image Intensifier (optional)
- Tri-mode 12 in/9 in/7 in image intensifier
- 12 in.: 1.2 lp/mm
- 9 in.: 1.4 lp/mm
- 7 in.: 2.0 lp/mm
- DQE: 65% (typical)
- Tri-mode 12 in/9 in/7 in image intensifier
Video Camera
- High resolution CCD camera.
- 525 line/frame.
- Full frame capture.
- 360° motorized rotation.
- Left-right image reversal.
- Top-bottom image reversal.
- Negative mode.
- Computer controlled features: (optional)
- Gain.
- Blanking.
- Camera iris.
Video Monitors
- Dual 17 in high resolution, anti-glare, anti-static.
- Ambient room light sensor.
- Multi-LED X-ray “ON” indicator.
Hardcopy Options
- Thermal printer. (optional)
- DICOM. (optional)
- Image Capture Device. (optional)
- Laser aiming device. (optional)
C-Arm Dimensions
- Height: 71 / 91 in. (1805 / 2305 mm)
- Width: 35 in. (880 mm)
- Length: 76 in. (1930 mm)
- Weight: 695 lbs 315 kg.
Workstation Dimensions
- Height: 69 in (1760 mm)
- Width: 32 in (800 mm)
- Depth: 37 in (925 mm)