Carefusion Alaris BD MedSystem III (2860, 2863, 2865)
24-Hour QuotesPricing up to 50% below the cost of new,
with service and warranty included. Download Specifications
Carefusion Alaris BD MedSystem III (2860, 2863, 2865) Features
The MedSystem III is intended for facilities that utilize infusion pumps for the delivery of fluids, medications, blood, and blood products using continuous or intermittent delivery through clinically acceptable routes of administration; such as intravenous (IV), intra-arterial (IA), subcutaneous, epidural, enteral, or irrigation of fluid spaces.
The MedSystem III is a multi-channel infusion pump intended to deliver multiple infusions to a single patient.
- Three independent fluid delivery systems in the space of one
- Compact size
- Easy to Setup and use, yet provides advanced features
- Accommodates assorted container types
- Multiple delivery methods
- Accurate delivery of a variety of fluids
- Uses administration sets that provide free-flow protection
- Six available Device Types with configurable parameters (maximum and minimum rates, maximum volumes, baseline, and maximum pressures, and air-in-line thresholds) to achieve specific clinical applications
- Display infusion status for rate, volume remaining, and volume infused.
- Infusions can be programmed to deliver at a specified rate or over a specified period.
- Secondary mode allows fluids and medications to be delivered at two different rates, sequentially.
- Dose Rate Calculator (DRC) feature performs volumetric rate and/or dose rate calculations.
- With DRC activated, the MedSystem III displays infusion status for rate, dosing regimen, and drug name.
- Communications Protocol allows clinical monitoring, MedSystem III configuration, and maintenance.
- Field Maintenance Software (FMS) is available for Biomed to configure, service, and troubleshoot the MedSystem III.
Carefusion Alaris BD MedSystem III (2860, 2063, 2865) Specifications
Dimensions
- Height: 7.875 in (20 cm)
- Width: 6 in (15.24 cm)
- Depth: 2.10 in (5.33 cm)
- Weight: 5.1 lb (2.3 kg) – including pole clamp
Occlusion Pressure
- 15 PSI except for controller Pressure device type which is 3ft. H20
Operating Temperature
- 50 – 104° F (10 – 40° C)
Transport/ Storage Temperature
- -4 to 131° F ( -20 to 55° C)
Rate Range
- 0.1 – 999 milliliter per hour (each channel)
Volume Rage
- 0.1 – 999 milliliter (each Channel)
KVO Range
- 0.1 – 20.0 milliliter per hour
Rate Accuracy
- 1.0 – 999 ml/hr ± 5% with a standard deviation of 1.96 under specified conditions<?li>
- 1.0 – 0.9 ml/hr ± 10% with a standard deviation of 1.96.
Batteries
- Main – rechargeable Nicid Battery Pack
- Memory Backup – Non-rechargeable lithium
Battery Charge
- A fully charged battery has a minimum of 6 hours running time with all channels running at 125 milliliters per hour and backlight usage of 2 minutes per hour.
- The main battery retains 80% of its capacity after 500 charging cycles and retains 90% of its capacity after 3 months of continuous AC charging.
AC Power Requirements
- Voltage: 90 VAC to 132 VAC
- Frequency 47 Hz to 63 Hz
Delivery
- The 2865 features a delivery rate range of 0.1 – 999 milliliters/hour for each channel and a KVO rate range of 0.1 – 20.0 milliliters/hour.
- 2863 features a delivery rate range of 0.1 – 999 milliliters/hour for each channel and a KVO rate range of 0.1 – 20.0 milliliters/hour.
- The 2860 features a delivery rate range of 0.1 – 999 milliliters/hour for each channel and a KVO rate range of 0.1 – 20.0 milliliters/hour.