Medfusion 2010i
24-Hour QuotesPricing up to 50% below the cost of new,
with service and warranty included. Download Specifications
Medfusion 2010i Features
The Medfusion 2010i is an ambulatory syringe pump fitting a variety of needs and applications. The 2010I Syringe Pump can be used when meticulous low-volume injection drugs are needed. The 2010i can be customized for specific applications, here you can choose from four different deliveries; Body Weight Mode, Mass Mode, Continuous Mode, and Volume Over Time Mode (dose-volume/delivery time). The syringe pump has the ability to store up to 64 user regiments.
- Flow/Rate Range: 0.1-379.0 ML/HR Dependent on Syringe Size Selected.
- Accuracy: Plus or Minus 3%
- Power Requirements: AC Adapter or internal battery; Model 2010i has an optional 220 Volt AC charger
- 4.5 in x 3 in x 7.5 in
- Syringe Size 1-60 ml
- Alarms: Near Empty, Empty, Volume Limit, Occlusion, Systems Malfunction, Low Battery, Depleted Battery, Syringe Not Loaded, Check Clutch, Invalid Syringe Size.
- Model: 2010 vs 2010i: 2010i has more conversion options for measurement
Medfusion 2010i Specifications
Dimensions
- Height: 3 in
- Width: 4.5 in
- Long: 7.5 in
- Weight: 2.5 lbs
General
- Flow rate: 0.01-378.0 ml/hr dependent upon syringe size
- Bolus rate: up to 378.0 ml/hr dependent upon syringe size
- Total Volume Delivered: 0.0000-9999 mg. ml or units
- Accuracy: +/-3%
- Infusion Modes: Body Weight, Mass, Continuous and Volume/time
- Accepts B-D, Monoject, and Terumo syringes
- Accepts: B-D 1 TB, 1, 3, 5, 10, 20, 30, 60 ml Luer lock
- Monoject 1 TB, 3, 6, 12, 20, 35, 60 ml Luer lock
- Terumo 1 TB, 3, 5, 10, 20, 30, 60 ml Luer lock
- B-D (glass) 1, 3, 5, 10 ml
- Automatically adjusts the output PSI for different syringe sizes
- Automatically sense syringe size, syringe barrel placement, and syringe plunger placement
- All syringes will fill to their maximum
- Optional 220-volt AC charger makes this syringe usable throughout the world
Power
- Power: AC-internal battery
- Recharge Time: No longer than 16 hours with the pump turned off
- Battery Capacity: at 25C a 16-hour charge will operate the pump for approximately 10 hours at 5.0 ml/hr with a 60 ml syringe
- At 25 degrees C, a 16-hour charge will pump approx. 10 hrs at 5.0 ml/hr with a 60 ml syringe
Infusion Modes
Body Weight Mode
- µg/kg/min µg/kg/min
- MG/kg/min MG/kg/HR
Mass Modes
- µg/min µg/HR
- MG/min MG/HR
- mU/min mU/HR
- U/min U/HR
Continuous Mode
- ml/min ml/HR
Volume Over Time Mode
- Volume/time
Flow Rate
- 0.01 to 378.0 ml/HR dependent on the syringe size selected
Bolus Delivery
- In Body Weight, Mass, and Continuous Modes
Alarms/ Alerts
- Near Empty (5 to 60 minutes), Empty, Dose Volume Delivered, Bolus
Complete, Bolus Delivery, Occlusion, System Malfunction, Low Battery, Depleted
Battery, Syringe Pops Out, Invalid Size, Invalid Number, Check - Clutch, Alarm Temporary Delay: 2 or 60 minutes, Alarm Audio Volume: Soft or Loud,
Load Syringe Plunger
Status Alerts Visual, Non-Audible
- Stop/Program, Deliver, Battery In Use, Battery Charging, Battery Depleted,
Priming, Standby Mode, Back Light
Total Delivery
- From 0.0000 to 9999.9999 MG
- From 0.00 to 9999.99 ml
- From 0.0000 to 9999.9999 U
Syringe Options/Sizes
- Becton-Dickinson (B-D*): 1,3,5,10,20,30,60 ml
- Monoject* (Mono): 1,3,6,12,20,35,60 ml
- Terum (Teru): 1,3,5,10,20,30,60 ml
- Becton-Dickinson Glass (BG-G) 1,3,5,10 ml
- Abboject* *Abot): 50 ml
Refurbished The Medfusion 2010i Infusion Pump
Soma Tech Intl offers the Medfusion 2010i Infusion Pumps up to 50% below OEM prices with the same service and warranty as new. The Medfusion 2010i is not only technically refurbished but also cosmetically refurbished to make it work and look like new. When the refurbishing process starts on the Medfusion 2010i, it is carefully inspected and tested by our own in-house biomedical engineers who are highly skilled and certified to work on infusion pumps. If necessary, parts are replaced to make sure that all aspects of the Medfusion 2010i are to the original engineering manufacturer’s specifications. This ensures that the unit is working the same way that it originally left the manufacturer. After the Medfusion 2010i is working like new it undergoes a special cosmetic restoration process. The Medfusion 2010i is cleaned, minor scratches and dents are repaired, it’s painted, and new decals are replaced if necessary. After this strenuous refurbishing process, the unit functions and looks like new. Soma Tech Intl is ISO 13485 certified and ensures that the best quality is given to all capital medical equipment that they sell. Before the Medfusion 2010i is labeled patient-ready and packed up for shipment, it goes through one last quality control test that makes sure all aspects are again working as they should. If you are looking to buy a refurbished Medfusion 2010i infusion pump contact soma at [email protected] or call us at 1-800-GET-SOMA.
Soma Tech Intl is actively looking to purchase a Medfusion 2010i If you or your facility is looking to sell your Medfusion 2010i, be sure to contact Soma Tech Intl. Soma has a dedicated and experienced purchasing department that helps you through the entire process. Selling your used or pre-owned Medfusion 2010i takes out the middle man in the equation. This also ensures that your medical facility gets top dollar for your equipment. Soma is 100% HIPPA compliant and makes sure that your equipment gets to our facility in the same condition that it left yours. Soma Tech Intl also gives the option for trade-in credit on your medical equipment. If you have medical equipment you would like to sell to Soma, send us a message here!